Skip to content

ATTAIN STARFIX® 419578Recalls & reported problems

Medtronic, Inc. · LEAD 419578 DEPLOYABLE 78CM EN, US · Class 3 · Implant

Quick summary

This is a Medtronic Attain Starfix 419578 deployable pacing lead, a wire implanted in the heart for permanent pacing. The 32 reports mostly describe the lead no longer capturing the heartbeat properly or needing unusually high pacing thresholds, along with fractured leads, high impedance, and leads that moved out of position. Some patients reported muscle or nerve stimulation, and a few had infections. 29 reports describe injuries and no deaths. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
32
Deaths reported
0
Injuries reported
29
Recalls naming it
0

Summary

The FDA's MAUDE database contains 32 adverse event reports that list this exact device (DI 00885074511986), received between Feb 9, 2017 and Dec 12, 2025 — classified as 29 injuries, 3 malfunctions.

The most frequently reported specific device problem is “High Capture Threshold” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Capture Threshold13
  • Adverse Event Without Identified Device or Use Problem9
  • Capturing Problem5
  • Fracture5
  • Failure to Capture4
  • High impedance4
  • Appropriate Device Problem Term/Code Not Available2
  • Unstable Capture Threshold2

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • No Known Impact Or Consequence To Patient8
  • Undesired Nerve Stimulation4
  • Unspecified Infection4
  • Muscle Stimulation3
  • Bacteremia1
  • Cardiac Perforation1
  • Cardiac Tamponade1

Recent adverse event reports

The 25 most recent of 32 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 12, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD DISLODGED. THE LEAD WAS EXPLANTED WITH NO REPLACEMENT. IT WAS ALSO REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD HAD FRACTURED. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23790756 ↗
  • InjuryAug 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFECTION WITH BACTEREMIA. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED. DURING EXPLANT THE LEFT VENTRICULAR (LV) LEAD, IT WAS NOTED TO HAVE TORE ¿A SIGNIFICANT HOLE IN THE CORONARY SINUS". A POSTERIOR EFFUSION WAS SEEN AND PATIENT REQUIRED A STERNOTOMY. IT WAS NOTED THAT THE PATIENT IS RECOVERING AS EXPECTED. THE PATIENT IS A PARTICIPANT IN A CLINICAL STUDY. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bacteremia; Cardiac Perforation

    MAUDE report 22834158 ↗
  • InjuryOct 17, 2024Failure to Capture; High impedance; Failure to Capture; High impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED NOISE, OUT OF RANGE (OOR) HIGH RV COIL AND SUPERIOR VENA CAVA (SVC) COIL IMPEDANCE WARNINGS, IN ADDITION TO SUSPECTED HIGH VOLTAGE RV COIL AND SVC COIL FRACTURES. ADDITIONALLY, IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD EXHIBITED OOR HIGH IMPEDANCE AND NO CAPTURE. IT WAS NOTED THAT ALARMS AND DETECTIONS WERE TURNED OFF. THE RV LEAD AND LV LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20471086 ↗
  • InjurySep 30, 2024High Capture Threshold; Unstable Capture Threshold; High Capture Threshold; Unstable Capture Threshold

    IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD EXHIBITED ELEVATED AND VARYING PACING THRESHOLDS AND THE PATIENT WAS FEELING IT. MONITORING WAS DISABLED AND A FIXED OUTPUT WAS SET. THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 20333819 ↗
  • InjuryAug 5, 2024High Capture Threshold; High Capture Threshold

    IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH THRESHOLD MEASUREMENTS. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19906853 ↗
  • InjuryJun 5, 2024High Capture Threshold; High Capture Threshold

    IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH THRESHOLDS AND LASER LEAD EXTRACTION WAS ATTEMPTED. IT WAS NOTED THAT THERE WAS AN APPARENT HOLE OBSERVED IN THE PATIENT'S HEART AND THERE WAS EXTRACTION TAMPONADE. THE LV LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Cardiac Tamponade

    MAUDE report 19467148 ↗
  • MalfunctionApr 11, 2024Low impedance; Low impedance

    IT WAS REPORTED THAT AN ALERT TRIGGERED FOR LOW PACING IMPEDANCE ON THE LEFT VENTRICULAR (LV) LEAD. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19089979 ↗
  • InjuryMay 17, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD, RIGHT VENTRICULAR (RV) LEAD AND CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WERE NOT FUNCTIONING WITHIN NORMAL TOLERANCES. THE ENTIRE CRT-D SYSTEM WAS EXPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16947895 ↗
  • InjuryMar 3, 2023Break; Break

    IT WAS REPORTED THAT DURING THE IMPLANT OF A NEW DEVICE, IT WAS OBSERVED THAT THERE WAS AN INSULATION BREACH ON THE LEFT VENTRICULAR (LV) LEAD. THE LEAD WAS REPAIRED AND THE LEAD WAS TESTED. THE LEAD THRESHOLD AND IMPEDANCE MEASUREMENTS WERE STABLE AND THE LV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16479291 ↗
  • MalfunctionNov 7, 2022Capturing Problem; Capturing Problem

    IT WAS FURTHER REPORTED VIA FOLLOW-UP COMMUNICATION THAT THE POSSIBLE PREMATURE DEPLETION WERE LIKELY DUE TO THE INCREASING RV AND LV THRESHOLDS. THE LV LEAD WAS REPROGRAMMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15747956 ↗
  • InjuryAug 3, 2022Failure to Capture; Fracture; High impedance; Failure to Capture; Fracture; High impedance

    IT WAS REPORTED THAT LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH AND UNDEFINED IMPEDANCE, NO CAPTURE, AND A POSSIBLE FRACTURE. THE PROXIMAL PORTION OF THE LV LEAD WAS CUT AND CAPPED AND A NEW LV LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15156986 ↗
  • InjuryMay 26, 2022High Capture Threshold; High Capture Threshold

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED UNEXPECTED LONGEVITY BASED ON PRIOR LONGEVITY ESTIMATES. HIGH OUTPUTS WERE NOTED ON THE LEFT VENTRICULAR (LV) LEAD, AND OUTPUTS WERE NOTED TO HAVE BEEN PROGRAMMED UPWARD. THE CRT-D AND LV LEAD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14518455 ↗
  • InjuryJan 13, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFECTION. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 13252247 ↗
  • InjuryNov 10, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING NON-DEVICE RELATED CARDIAC PROCEDURES, THE PATIENT EXPERIENCED PAINFUL, INTERMITTENT, AND VISIBLE PHRENIC NERVE STIMULATION FROM THE LEFT VENTRICULAR (LV) LEAD THAT DID NOT RESOLVE. THE PHYSICIAN ATTEMPTED TO REPROGRAM THE LV LEAD TO NO AVAIL. THE LV LEAD WAS TURNED OFF AND REMAINS IN THE PATIENT. THE PATIENT IS A PARTICIPANT IN A CLINICAL STUDY. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Pain; Undesired Nerve Stimulation

    MAUDE report 12791665 ↗
  • InjuryJul 15, 2021Failure to Capture; High Capture Threshold; Failure to Capture; High Capture Threshold

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) REACHED POSSIBLE EARLY BATTERY DEPLETION. THE LEFT VENTRICULAR (LV) LEAD HAD HIGH THRESHOLD WITH NO CAPTURE AT MAXIMUM OUTPUT. THE LV LEAD AND THE DEVICE WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12172996 ↗
  • InjuryAug 10, 2020High Capture Threshold; High Capture Threshold

    IT WAS REPORTED THAT THERE WAS HIGH THRESHOLD ON THE LEFT VENTRICULAR (LV) LEAD. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10387353 ↗
  • InjuryMay 7, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXTRACTED DUE TO A STAPHYLOCOCCUS INFECTION. A SINGLE CHAMBER PACEMAKER AND RIGHT VENTRICULAR (RV) LEAD WERE IMPLANTED AS THE PATIENT IS PACEMAKER DEPENDENT. A NEW CRT-D SYSTEM WAS IMPLANTED ONE WEEK LATER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 10033493 ↗
  • InjuryOct 29, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS EXPERIENCING PHRENIC NERVE AND DIAPHRAGMATIC STIMULATION. THE LEFT VENTRICULAR (LV) LEAD WAS CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Device Overstimulation of Tissue; Muscle Stimulation; Undesired Nerve Stimulation

    MAUDE report 9250840 ↗
  • InjuryOct 16, 2019Adverse Event Without Identified Device or Use Problem; High Capture Threshold; Adverse Event Without Identified Device or Use Problem; High Capture Threshold

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED PHRENIC NERVE STIMULATION. ADDITIONAL INFORMATION REPORTED THE LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH THRESHOLDS. THE LV LEAD WAS CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Undesired Nerve Stimulation

    MAUDE report 9196687 ↗
  • MalfunctionOct 1, 2019Fracture; Fracture

    THE LEFT VENTRICULAR (LV) LEAD WAS RETURNED TO THE MANUFACTURER WITH NO INFORMATION AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9141289 ↗
  • InjuryAug 14, 2019Fracture; High impedance; Fracture; High impedance

    IT WAS REPORTED THAT THERE WAS HIGH IMPEDANCE AND POSSIBLE PARTIAL FRACTURE ON THE LEFT VENTRICULAR (LV) LEAD. THE LEAD WAS INACTIVATED AND REMAINS IN THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8892936 ↗
  • InjuryMar 28, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY COMPETITOR THAT THE LEFT VENTRICULAR LEAD WAS EXPLANTED DUE TO AN INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 8462079 ↗
  • InjurySep 6, 2018High Capture Threshold; High Capture Threshold

    IT WAS REPORTED THAT THE PATIENT'S RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCE. IT WAS SUSPECTED THE RV LEAD HAD FRACTURED. THE RV LEAD WAS EXPLANTED AND REPLACED. IT WAS ALSO REPORTED THE PATIENT'S LEFT VENTRICULAR (LV) LEAD EXHIBITED HIGH THRESHOLDS. THE LV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7851572 ↗
  • InjuryJul 9, 2018Premature Elective Replacement Indicator; Battery Problem; Capturing Problem; High Capture Threshold; Capturing Problem

    IT WAS REPORTED THE LEFT VENTRICULAR (LV) LEAD DISPLAYED INCREASED THRESHOLDS WHICH AFFECTED THE BATTERY LIFE OF THE CARDIAC RESYNCH RONIZATION THERAPY DEFIBRILLATOR (CRT-D). THE LV LEAD WAS INACTIVATED HOWEVER THE PATIENT EXPERIENCED DYSPNEA ON EXERTION AND THE LEAD WAS TURNED BACK ON. IT WAS FURTHER REPORTED THERE WAS CHRONIC LEFT DIAPHRAGMATIC STIMULATION. AS THE STIMULATION DID NOT BOTHER THE PATIENT, ONLY THE GENERATOR CHANGE WAS PERFORMED DUE TO ELECTIVE REPLACEMENT INDICATOR (ERI). THE LEAD REMAINS IN USE AND THE DEVICE WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Dyspnea; Muscle Stimulation

    MAUDE report 7671016 ↗
  • InjuryJun 5, 2018High Capture Threshold; High Capture Threshold

    IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD EXHIBITED A HISTORY OF HIGH THRESHOLDS AND WAS NOT USABLE. THE LV LEAD WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7571716 ↗

Other permanent pacemaker Electrode devices

All permanent pacemaker Electrode devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.