Skip to content

DeVilbiss 7314D-DRecalls & reported problems

Devilbiss Healthcare LLC · Vacu-Aide QSU, North America · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885304000587), received between Mar 23, 2023 and Mar 23, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Suction Problem” (1 report).

3 other recalls cover devices cleared under the same FDA submission (K982304); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Suction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 23, 2023Suction Problem; Suction Problem

    DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A SUCTION UNIT BY A PROVIDER, WHO STATED THAT THE UNIT HAS "WILL NOT SUCTION." THERE WAS NO REPORT OR EVIDENCE OF ILLNESS, INJURY OR MEDICAL TREATMENT ASSOCIATED WITH THE COMPLAINT. THE DEVICE WAS RETURNED TO DEVILBISS FOR EVALUATION, WHERE IT WAS DETERMINED THE UNIT WAS UNABLE TO PRODUCE VACUUM PRESSURE TO THE MINIMUM SPECIFICATION. THE ROOT CAUSE WAS DETERMINED TO BE AN UMBRELLA-STYLE CHECK VALVE THAT HAD DISLODGED FROM THE COMPRESSOR CYLINDER AND A LOOSE MANIFOLD. DEVILBISS HAS AN ACTIVE CAPA ASSOCIATED WITH VALVE/CYLINDER COMPLAINTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16602968 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K982304. One submission often covers many models, so check the product and lot details in each record.

  • Z-2246-2016 · May 23, 2016 · Terminated
    DeVilbiss Model 7305P-D Powered Suction Unit
  • Z-2319-2013 · Mar 7, 2013 · Terminated
    DeVilbiss Vacu-Aide QSU Suction Unit (Disposable Suction Container/Filter), 800 ML DISP Container, Unassembled, 800 ML DISP Container w/6' patient tubing, and 800 ML DISP Container
  • Z-0863-2009 · Oct 27, 2008 · Terminated
    DeVilbiss Model 7310PR-D VacuAide Compact Suction Unit. A portable, AC-powered or compressed air-powered suction pump. The device may include a microbial filter.

Other Apparatus, Suction, Ward Use, Portable, Ac-Powered devices

All Apparatus, Suction, Ward Use, Portable, Ac-Powered devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.