Drive DeVilbiss Vacu-Aide Suction Unit 7325P-DRecalls & reported problems
Drive Devilbiss Healthcare · Vacu-Aide w/battery, North America · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00885304032564), received between Sep 19, 2022 and Jun 15, 2023 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Suction Problem” (3 reports).
3 other recalls cover devices cleared under the same FDA submission (K982304); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Suction Problem3
- Suction Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 15, 2023Suction Problem; Suction Problem
DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A SUCTION UNIT BY A PROVIDER, WHO STATED THE "UNIT WILL NOT SUCTION." THERE WAS NO REPORT OR EVIDENCE OF ILLNESS, INJURY OR MEDICAL TREATMENT ASSOCIATED WITH THE COMPLAINT. THE DEVICE WAS RETURNED TO DEVILBISS FOR EVALUATION, WHERE IT WAS DETERMINED THE UNIT WAS UNABLE TO PRODUCE VACUUM PRESSURE TO SPECIFICATION. THE ROOT CAUSE WAS THE UMBRELLA-STYLE CHECK VALVE THAT DISLODGED FROM THE COMPRESSOR CYLINDER AND A LOOSE MANIFOLD. DEVILBISS ADDRESSED CHECK VALVE COMPLAINTS VIA CAPA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17134912 ↗MalfunctionJun 15, 2023Suction Problem; Suction Problem
DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A SUCTION UNIT BY A PROVIDER, WHO STATED THE "[UNIT HAS] VERY LOW SUCTION." THERE WAS NO REPORT OR EVIDENCE OF ILLNESS, INJURY OR MEDICAL TREATMENT ASSOCIATED WITH THE COMPLAINT. THE DEVICE WAS RETURNED TO DEVILBISS FOR EVALUATION, WHERE IT WAS DETERMINED THE UNIT WAS UNABLE TO PRODUCE VACUUM PRESSURE TO SPECIFICATION. THE ROOT CAUSE WAS THE UMBRELLA-STYLE CHECK VALVE THAT DISLODGED FROM THE COMPRESSOR CYLINDER AND A LOOSE MANIFOLD. DEVILBISS ADDRESSED CHECK VALVE COMPLAINTS VIA CAPA.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17134098 ↗MalfunctionApr 14, 2023Suction Problem; Suction Problem
DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A SUCTION UNIT BY A PROVIDER, WHO STATED THAT THE "UNIT HAS NO SUCTION." THERE WAS NO REPORT OR EVIDENCE OF ILLNESS, INJURY OR MEDICAL TREATMENT ASSOCIATED WITH THE COMPLAINT. THE DEVICE WAS RETURNED TO DEVILBISS FOR EVALUATION, WHERE IT WAS DETERMINED THE UNIT WAS UNABLE TO PRODUCE VACUUM PRESSURE TO THE MINIMUM SPECIFICATION. THE ROOT CAUSE WAS THE UMBRELLA-STYLE CHECK VALVE THAT DISLODGED FROM THE COMPRESSOR CYLINDER. DEVILBISS HAS A CAPA FOR THE CHECK VALVE COMPLAINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16742985 ↗MalfunctionSep 19, 2022Suction Failure; Suction Failure
DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A SUCTION UNIT BY A PROVIDER, WHO REPORTED THAT THE UNIT "QUIT SUCTIONING, AND THE BATTERY LIGHT IS ON, HOWEVER, THERE IS NO SUCTIONING AND NO SOUNDS." THERE WAS NO REPORT OR EVIDENCE OF ILLNESS, INJURY OR MEDICAL TREATMENT ASSOCIATED WITH THE COMPLAINT. THE UNIT WAS DISCARDED AND COULD NOT BE EVALUATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15445787 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K982304. One submission often covers many models, so check the product and lot details in each record.
- Z-2246-2016 · May 23, 2016 · TerminatedDeVilbiss Model 7305P-D Powered Suction Unit
- Z-2319-2013 · Mar 7, 2013 · TerminatedDeVilbiss Vacu-Aide QSU Suction Unit (Disposable Suction Container/Filter), 800 ML DISP Container, Unassembled, 800 ML DISP Container w/6' patient tubing, and 800 ML DISP Container
- Z-0863-2009 · Oct 27, 2008 · TerminatedDeVilbiss Model 7310PR-D VacuAide Compact Suction Unit. A portable, AC-powered or compressed air-powered suction pump. The device may include a microbial filter.
Other Apparatus, Suction, Ward Use, Portable, Ac-Powered devices
- Penumbra System PMX110Penumbra, Inc. · APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED166 reports
- IntelliCart™ 00514010400Zimmer Orthopaedic Surgical Products, Inc. · APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED112 reports
- Penumbra System PMX220Penumbra, Inc. · APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED102 reports
- DeVilbiss 7314P-DDevilbiss Healthcare LLC · Apparatus, Suction, Ward Use, Portable, Ac-Powered69 reports
- PIRANHA 2208011Richard Wolf GmbH · APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED54 reports
- Neptune 0770100000Stryker Corporation · APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED1 recall0 reports
- LCSU 4, 800 ml, RTCA Complete Unit 880052Laerdal Medical AS · Apparatus, Suction, Ward Use, Portable, Ac-Powered1 recall0 reports
- Neptune 0711001000Stryker Corporation · APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page