TRUCLEARTM ULTRA Plus Device 72203012Recalls & reported problems
Smith & Nephew, Inc. · TRUCLEAR ULTRA RECIPROCATING MORC. 4.0 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00885554027617), received between Apr 13, 2018 and Oct 19, 2020 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break2
- Adverse Event Without Identified Device or Use Problem1
- Appropriate Device Problem Term/Code Not Available1
- Bent1
- Detachment Of Device Component1
- Detachment of Device or Device Component1
- Material Fragmentation1
- Mechanical Jam1
Patient problems
- Device Embedded In Tissue or Plaque1
- No Code Available1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
- Radiation Exposure, Unintended1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryOct 19, 2020Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Mechanical Jam; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE PRODUCT DID NOT PERFORM ITS RECIPROCATING FUNCTION TO BE ABLE TO MAKE A RECESSION OF THE INTRA-UTERINE MYOMA. NO OTHER INSTRUMENT WAS USE AND THE PRODUCT WAS REMOVED FROM THE HAND PIECE. THE CASE WAS ABORTED, AND WAS RESCHEDULED ON (B)(6) 2020.
Patient problems: No Code Available
MAUDE report 10696888 ↗InjuryAug 1, 2018Material Fragmentation; Detachment of Device or Device Component; Appropriate Device Problem Term/Code Not Available; Material Fragmentation
ACCORDING TO THE REPORTER, WHEN THE DEVICE WAS BEING APPLIED TO THE UTERUS DURING A HYSTEROSCOPIC MYOMECTOMY, THE INNER SHAFT OF THE DEVICE SPUN OUT AND WRAPPED AROUND THE OUTER TUBE OF THE DEVICE. IT WAS VERY SHARP AND JAGGED METAL. THIS RESULTED TO AN EXTENSION WITH THE SURGICAL TIME. NOTHING WAS DISENGAGED FROM THE HANDPIECE. IT WAS REQUIRED TO OPEN ANOTHER DEVICE AND AN X-RAY TO ENSURE NO MATERIAL FROM THE DEVICE WAS LEFT BEHIND IN CAVITY. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE AND THE PATIENT HAD NO POST OPERATIVE ISSUES. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient; Radiation Exposure, Unintended
MAUDE report 7739253 ↗InjuryMay 16, 2018Break; Break
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE PIECE OF THE UNIT BROKE INTO THE PATIENT. BROKEN PIECES WAS LOCATED THROUGH HYSTEROSCOPY AND IT WAS RETRIEVED VIA GRASPERS. THE PROCEDURE WAS EXTENDED AND IT RAN ABOUT 2 AND A HALF HOURS.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 7520619 ↗MalfunctionApr 13, 2018Bent; Break; Detachment Of Device Component; Bent; Break; Detachment Of Device Component
ACCORDING TO THE REPORTER, DURING THE PROCEDURE, THE DEVICE BROKE WHILE RESECTING A 4CM SUBMUCOUS FIBROID. THE DEVICE STARTED JAMMING UP AND THEN IT APPEARED THAT THE RECIPROCATING BLADE WAS NO LONGER COMPLETELY WITHIN THE SHAFT. THE INSTRUMENT WAS REMOVED FROM THE PATIENT AND INSPECTED AND APPEARED BENT. THEY USED A SECOND DEVICE AND DURING THE PROCEDURE NOTICED A SMALL PIECE OF METAL INSIDE THE PATIENT'S UTERINE CAVITY. THE SURGEON MANUALLY ROTATED THE BROKEN DEVICE INTO AN OPEN POSITION TO USE IT FOR SUCTION AND PLACED IT BACK THROUGH THE HYSTEROSCOPE AND PLACED IT DIRECTLY OVER THE METAL FRAGMENT AND TURNED ON THE HANDPIECE SUCTION TO GET IT OUT OF THE CAVITY. THERE WAS NO PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7428972 ↗
Other HYSTEROSCOPE (AND ACCESSORIES) devices
- Olympus WA2T430AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)783 reports
- Olympus WA2T412AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)453 reports
- Olympus A22001AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)366 reports
- Olympus WA2T470AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)323 reports
- Olympus A22002AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)297 reports
- Olympus A22003AOLYMPUS Winter & Ibe GmbH · Hysteroscope (And Accessories)198 reports
- n.a. 27040XAKarl Storz GmbH & Co. KG · HYSTEROSCOPE (AND ACCESSORIES)102 reports
- Myosure Reach Tissue Removal Device 10-401FCHologic, Inc. · HYSTEROSCOPE (AND ACCESSORIES)4 deaths99 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page