Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885556132975), received between Apr 20, 2021 and Nov 7, 2023 — classified as 3 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K092748, K210980); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Mar 18, 2025Open, Classified
Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.
Z-1806-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Break2
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryNov 7, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING A META-TAN SCREW PLANNED REMOVAL, THE BACK END OF THE META-TAN LAG/COMPRESSION SCREW 90MM/85MM FRACTURED OFF. AFTER A DELAY OF MORE THAN 30 MIN, THE SCREW COULD NOT BE REMOVED, IT REMAINS IN THE PATIENT. NO ADDITIONAL INTERVENTION IS PLANNED TO REMOVE THE SCREW FROM HIS BODY. PATIENT'S CURRENT HEALTH STATUS IS GOOD.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 18084706 ↗InjuryAug 17, 2022Break; Break
IT WAS REPORTED THAT, DURING SURGERY OF NAIL REMOVAL (PLANNED SURGERY), WHEN REMOVING'S THE LAG SCREW, THE TIP OF THE TWO (2) META-TAN LAG SCREW DRIVER WERE BROKEN. USING OTHER TOOL SCREW EXTRACTOR SIZE 1, THE SURGERY WAS COMPLETED, BUT ALSO IT WAS BROKEN WHEN IT WAS REMOVED FROM THE EXPLANTED LAG SCREW. DEVICE INSIDE THE PATIENT. PIECES FELL INSIDE THE PATIENT AND IT IS UNKNOWN IF/HOW THE PIECES WERE EXTRACTED. THE PROCEDURE WAS COMPLETED USING A S+N BACK-UP DEVICE. SURGERY WAS 30 MIN DELAYED. PATIENT WAS NOT HARMED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15245681 ↗InjuryApr 20, 2021Break; Break
IT WAS REPORTED THAT, DURING A PLANNED SURGERY OF META-TAN REMOVAL, IT WAS DIFFICULT TO REMOVE THE META-TAN NAIL WITH AN UNKNOWN META-TAN LAG SCREW DRIVER ((B)(4)). THEREFORE, META-TAN LAG SCREW DRIVER WAS CHANGED TO EASY OUT SMALL . THE NAIL WAS REMOVED, BUT PART OF AN UNKNOWN LAG SCREW ((B)(4)) AND THE EASY OUT SMALL ((B)(4)) BROKE INSIDE THE PATIENT. ALL PIECES WERE REMOVED. THE PROCEDURE WAS COMPLETED WITH A LESS THAN 30 MINS DELAY USING THE SAME DEVICE. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11696358 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K092748, K210980. One submission often covers many models, so check the product and lot details in each record.
- Z-2456-2026 · Mar 16, 2026 · Open, ClassifiedTRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816
- Z-2443-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 36CM 130 DEGREE LEFT, REF 71647336, Femur nail
- Z-2444-2023 · Jul 24, 2023 · Open, ClassifiedTRIGEN TROCHANTERIC ANTEGRADE NAIL LIME 11.5MM X 40CM 130 DEGREE LEFT, REF 71647340, Femur nail
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE Intramedullary Limb Lengthening System LSB5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page