Skip to content

TRIGEN recallZ-1806-2025 · Smith & Nephew, Inc. · initiated Mar 18, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smith & Nephew, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Integrated Lag/Compression Screw Kit has a compression screw with hexagon socket that may be undersized, which causes the screw to be non-functional, which would require replacement with another adequate size screw, and exchange may exceed 30 minutes.

FDA root cause category: Process control

What the company is doing

On 3/18/2025, recall notices were mailed and emailed to customers who were asked to do the following: 1) Inspect your inventory and locate any affected devices and quarantine them. 2) If you are a sales rep, district office, or distributor, you must notify your customers of the field action and ensure that the required actions are complete. 3) Complete and return the notice response section via email to [email protected] If you have any questions email the firm at the email address above.

Recall details

Recall number
Z-1806-2025
Status
Open, Classified
Initiated by the company
Mar 18, 2025
Posted by the FDA
May 19, 2025
Recalling firm
Smith & Nephew, Inc.
Quantity
75
Product code
HSB
FDA submissions
K092748, K210980
Distribution
Worldwide - US Nationwide distribution in the states of FL, NH, TX, CA, OH, NY, AL, IA, MA, NV, MO, WA, LA, TN, NC, KY, NM, NJ, AZ and the country of ZA.

Product description

META-TAN LAG/COMPRESSION SCREW KIT 90MM/85MM, REF: 71642690

Code information (models, lots, serial numbers)

UDI-DI: 00885556132975. Lot: 24CT81426. Expiration: 25-Mar-2034

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

149 devices cleared under K092748, K210980 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page