Skip to content

TRIGEN 71631418Recalls & reported problems

Smith & Nephew, Inc. · META-TAN LAG SCREW DRILL SLEEVE · Class 2

All models of TRIGEN (381)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885556158425), received between Feb 24, 2026 and Feb 24, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Failure to Align” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Align1
  • Unstable1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2026Unstable; Failure to Align

    IT WAS REPORTED THAT, DURING A THR SURGERY, A META-TAN LAG SCREW DRILL SLEEVE WAS LOOSE WITHIN THE META-TAN DROP. THE LOOSENESS CAUSED TO MUCH VARIANCE DURING THE GUIDE PIN INSERTION, WHICH ALLOWED THE GUIDE PIN TO MISS THE TARGETED NAIL. THE GUIDE PIN SHOT ANTERIOR TO THE NAIL APERTURE BUT ALSO WAS ABLE TO BE MANIPULATED TO HIT POSTERIOR AS WELL. THE PROCEDURE WAS SUCCESSFULLY RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING THE SAME DEVICE. THE CURRENT STATUS OF THE PATIENT IS UNKNOWN, BUT NO OTHER COMPLICATIONS WERE REPORTED, AND NO ADDITIONAL MEDICAL/SURGICAL INTERVENTION WAS NEEDED DUE TO THE INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24434126 ↗

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page