VLP MINI-MOD 74412417Recalls & reported problems
Smith & Nephew, Inc. · VLP TITANIUM 2.4MM X 17MM LOCKING SCREW T7 SELF-TAPPING · Class 2 · Implant
All models of VLP MINI-MOD (22)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885556348697), received between Aug 11, 2022 and Aug 11, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Positioning Failure” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K132886, K210837, K213126); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Positioning Failure1
Patient problems
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryAug 11, 2022Positioning Failure; Positioning Failure
IT WAS REPORTED THAT DURING A RIGHT 3H STANDARD PLATE IN BILATERAL WRIST CASE, THE 17MM LOCKING SCREW (VLP TI 2.4MM X 17MM LCK SCR T7 S-T NS) WENT THROUGH THE PLATE (EVOS VOLAR PLATE 3H RIGHT STD TI 48MM) IN THE DISTAL MOST HOLE ON THE ULNAR SIDE. THE PHYSICIAN WAS UNABLE TO BACK IT OUT AND ALREADY HAD OTHER SCREWS IN THE DISTAL FRAGMENT SO COULDN'T REMOVE THE PLATE TO REMOVE THE SCREW. SURGEON ISN'T REALLY CONCERNED BUT IT NEEDS TO BE REPORTED. THE SURGEON ASSUMED THAT THE TABS INSIDE THE HOLE BENT WHILE INSERTING THE SCREW AND DID NOT LOCK ONTO THE SCREW ALLOWING IT TO PASS THROUGH. THE SURGERY FINISHED WITH THE SAME DEVICE. NO INJURY OR SIGNIFICANT DELAY REPORTED. OPERATIVE REPORTS ARE NOT AVAILABLE.
Patient problems: Foreign Body In Patient
MAUDE report 15214391 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K132886, K210837, K213126. One submission often covers many models, so check the product and lot details in each record.
- Z-0396-2018 · Dec 11, 2017 · Terminatedsmith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Plating system, 2.0MM STRAIGHT PLATE, 8 HOLE VLP(R) MIN-MOD, REF 74442024, QTY: (
- Z-0397-2018 · Dec 11, 2017 · Terminatedsmith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Plating system, 2.0MM STOUT STRAIGHT PLATE, 8 HOLE VLP(R) MIN-MOD, REF 74442030,
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.