VLP MINI-MOD 74412418Recalls & reported problems
Smith & Nephew, Inc. · VLP TITANIUM 2.4MM X 18MM LOCKING SCREW T7 SELF-TAPPING · Class 2 · Implant
All models of VLP MINI-MOD (22)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556348741), received between Nov 29, 2023 and Jan 22, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Positioning Failure” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K132886, K210837, K213126); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Positioning Failure2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 22, 2024Positioning Failure; Positioning Failure
IT WAS REPORTED THAT DURING DISTAL RADIUS FIXATION, A VLP TI 2.4MM X 18MM LCK SCR T7 S-T WENT STRAIGHT THROUGH SCREW HOLE ON THE PLATE, IT WOULDN¿T ENGAGE IN PLATE AND WENT STRAIGHT THROUGH TO BONE. SURGERY WAS RESUMED WITHOUT ANY DELAY WITH A S+N BACK-UP DEVICE. CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18552404 ↗InjuryNov 29, 2023Positioning Failure; Positioning Failure
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, WHEN A VLP TI 2.4MM X 18MM LCK SCR T7 S-T WAS INSERTED INTO THE DISTAL, SECOND LINE, ULNAR SIDE HOLE OF AN EVOS VOLAR PLATE 4H LEFT STD STD TI 56MM, THE SCREW AND THE PLATE DID NOT LOCK, SINCE THE SCREW KEPT ROTATING. A BACK-UP, SAME SIZE SCREW WAS TRIED, HOWEVER THE SAME ISSUE OCCURRED. THE SURGERY WAS COMPLETED, WITHOUT INSERTING A SCREW IN THE SCREW HOLE, AND LEAVING THE PLATE INSIDE THE PATIENT. THIS DECISION WAS MADE, BECAUSE THERE WAS NO SAME SIZE PLATE AVAILABLE, THE ISSUE OCCURRED WITH ONLY ONE SCREW HOLE, AND THE OTHERS HAD NO PROBLEM, AND THE SURGEON THOUGHT IT WOULD BE OF HIGHER RISK TO REMOVE AND CHANGE ALL THE INSERTED SCREWS. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18226690 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K132886, K210837, K213126. One submission often covers many models, so check the product and lot details in each record.
- Z-0396-2018 · Dec 11, 2017 · Terminatedsmith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Plating system, 2.0MM STRAIGHT PLATE, 8 HOLE VLP(R) MIN-MOD, REF 74442024, QTY: (
- Z-0397-2018 · Dec 11, 2017 · Terminatedsmith&nephew Mini-Fragment Plating System, VLP(R) Mini-Mod Small Bone Plating System, EVOS(R) Mini Plating system, 2.0MM STOUT STRAIGHT PLATE, 8 HOLE VLP(R) MIN-MOD, REF 74442030,
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page