Skip to content

EVOS 72412416Recalls & reported problems

Smith & Nephew, Inc. · EVOS 2.4MM X 16MM LOCKING SCREW T7 SELF-TAPPING · Class 2 · Implant

All models of EVOS (190)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885556430590), received between Jan 5, 2022 and Jan 5, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Incomplete or Inadequate Connection” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K140814, K210837, K213126); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Incomplete or Inadequate Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 5, 2022Incomplete or Inadequate Connection; Incomplete or Inadequate Connection

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY AN EVOS 2.4MM X 16MM LCK SCR T7 S-T DID NOT LOCK INTO THE EVOS VOLAR PLATE 3H R STD 48MM. SURGERY WAS COMPLETED, WITHOUT ANY DELAY, WITH A SN BACK-UP DEVICE. NO HARM TO THE PATIENT OR FURTHER COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13171464 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K140814, K210837, K213126. One submission often covers many models, so check the product and lot details in each record.

  • Z-2945-2020 · Aug 24, 2020 · Terminated
    EVOS Screws - Product Usage: is a part of the EVOS MINI plating system, indicated for fracture fixation, arthrodesis, reconstruction, replantation or reduction of small bones and s
  • Z-0408-2018 · Nov 29, 2017 · Terminated
    smith&nephew, 2.4 MM, 6 HOLE, EVOS, FLEX PLATE, SST Plate, REF CATALOG NUMBER 72442360N, NON-STERILE, Rx only, (01)00885556456446(10), (11)171116 Bone plates and bone screws from

Other PLATE, FIXATION, BONE devices

All PLATE, FIXATION, BONE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page