EVOS 72402440NRecalls & reported problems
Smith & Nephew, Inc. · EVOS 2.4MM X 40MM CORTEX SCREW T7 SELF-TAPPING · Class 2 · Implant
All models of EVOS (190)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885556445945), received between Apr 28, 2021 and Apr 28, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K140814, K210837, K213126); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- Foreign Body In Patient1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 28, 2021Break; Break
IT WAS REPORTED THAT, DURING A PROXIMAL TIBIAL OSTEOSYNTHESIS, ONCE THE BONE HAD BEEN DRILLED AND AFTER MEASURING THE DEPTH OF THE HOLE, THE EVOS 2.4MM X 40MM CTX SCR T7 S-T WAS INSERTED THROUGH ONE OF THE HOLES IN THE PLATE. THE SCREW ENTERED WITHOUT ANY INCONVENIENCE, THEN WITH THE SCREWDRIVER IT WAS PROCEEDED TO FINISH INSERTING AND SECURING THE SCREW; HOWEVER, IT WAS EVIDENT THAT THE SCREW HEAD BROKE, THUS LEAVING THE SCREW BODY INSIDE THE PLATE. IT IS UNKNOWN IF THE BROKEN PART IS STILL IN THE PATIENT OR IF IT WAS RETRIEVED FROM THE BONE. THERE WAS NO SIGNIFICANT DELAY AND A SMITH AND NEPHEW BACK UP WAS AVAILABLE TO COMPLETE THE PROCEDURE. NO OTHER COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient; Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11736161 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K140814, K210837, K213126. One submission often covers many models, so check the product and lot details in each record.
- Z-2945-2020 · Aug 24, 2020 · TerminatedEVOS Screws - Product Usage: is a part of the EVOS MINI plating system, indicated for fracture fixation, arthrodesis, reconstruction, replantation or reduction of small bones and s
- Z-0408-2018 · Nov 29, 2017 · Terminatedsmith&nephew, 2.4 MM, 6 HOLE, EVOS, FLEX PLATE, SST Plate, REF CATALOG NUMBER 72442360N, NON-STERILE, Rx only, (01)00885556456446(10), (11)171116 Bone plates and bone screws from
Other PLATE, FIXATION, BONE devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.