VLP MINI-MOD 74401510Recalls & reported problems
Smith & Nephew, Inc. · VLP TITANIUM 1.5MM X 10MM CORTEX SCREW T4 SELF-TAPPING · Class 2 · Implant
All models of VLP MINI-MOD (22)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00885556489345), received between Nov 12, 2018 and Dec 29, 2021 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break4
- Fracture2
Patient problems
- No Consequences Or Impact To Patient2
- No Information2
- Foreign Body In Patient1
- Injury1
- No Clinical Signs, Symptoms or Conditions1
- Perforation1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryDec 29, 2021Break; Break
IT WAS REPORTED THAT, DURING A TRAUMA SURGERY CONDUCTED ON (B)(6) 2021 IN ORDER TO REDUCE A METACARPAL FRACTURE, A VLP MINI-MOD 1.5MM STRAIGHT PLATE 8 HOLE AND VLP TITANIUM 1.5MM X 10MM CORTEX SCREWS WERE PLACED. HOWEVER, WHEN FIXING THE VLP TITANIUM 1.5MM X 10MM CORTEX SCREWS, TWO OF THEM BROKE INSIDE THE PATIENT, LEAVING THE SCREW BODY INSIDE THE PATIENT'S BONE. THE VLP MINI-MOD 1.5MM STRAIGHT PLATE 8 HOLE WAS REMOVED. THE SCREWS REMAINED INSIDE THE PATIENT WITHOUT GENERATING ANY AFFECTATION. SURGERY WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A VLP MINI-MOD 2.0MM STRAIGHT PLATE 8 HOLE, WITHOUT ANY FURTHER ISSUE. PATIENT WAS NOT INJURED BEYOND THE INCIDENT REPORTED, AND THEIR CURRENT HEALTH CONDITION IS STABLE.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions; Perforation
MAUDE report 13112935 ↗InjuryApr 13, 2020Break; Break
IT WAS REPORTED THAT METACARPAL OSTEOSYNTHESIS SURGERY WAS PERFORMED ON THE PATIENT'S LEFT HAND. DISTAL AND PROXIMAL LOCKINGS WERE MADE BY PLACING THE THIRD SCREW-LOCKED FROM PROXIMAL TO DISTAL ORDER. AN X-RAY PLATE WAS TAKEN INTRAOPERATIVELY AND IT WAS FOUND THAT THE VLP TI 1.5MM X 9MM CTX SCR T4 S-T PROXIMAL SCREW WAS BROKEN. DESPITE THIS, THE FRACTURE WAS ALIGNED. THE BROKEN SCREW WAS LEFT INSIDE THE PATIENT AND THE PROCEDURE WAS COMPLETED AS PLANNED.
Patient problems: Injury
MAUDE report 9953312 ↗InjuryJan 23, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING SURGERY WHEN INSERTING THE SCREWS, THE HEADS SHEARED OFF MID-CASE LEAVING JUST THE SCREW SHAFT IN THE BONE. BONE WAS NORMAL AND NO EXCESSIVE FORCE WAS USED. THE SHAFTS WERE LEFT IN THE BONE AS THEY WERE FLUSH WITH THE CORTICAL WALL AND WOULD CAUSE MORE DAMAGE TRYING TO REMOVE. THERE WAS A DELAY BETWEEN 0 AND 30 MIN. A BACK-UP DEVICE WAS NOT AVAILABLE, THE PROCEDURE WAS COMPLETED WITH THE DEVICE IN QUESTION.
Patient problems: No Information
MAUDE report 9621185 ↗InjuryJan 23, 2020Fracture; Fracture
IT WAS REPORTED THAT DURING SURGERY WHEN INSERTING THE SCREWS, THE HEADS SHEARED OFF MID-CASE LEAVING JUST THE SCREW SHAFT IN THE BONE. BONE WAS NORMAL AND NO EXCESSIVE FORCE WAS USED. THE SHAFTS WERE LEFT IN THE BONE AS THEY WERE FLUSH WITH THE CORTICAL WALL. THERE WAS NO FURTHER PATIENT IMPART AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A SLIGHT DELAY OF 30 MINUTES OR LESS.
Patient problems: No Information
MAUDE report 9621182 ↗MalfunctionNov 13, 2018Break; Break
IT WAS REPORTED THAT INTRAOPERATIVELY, WHEN THE SURGEON WAS PLACING THE SCREW ON THE PLATE, THE HEAD OF THE SCREW WAS BROKEN, LEAVING ONE PART ENCRUSTED AND THE OTHER OUTSIDE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8063909 ↗MalfunctionNov 12, 2018Break; Break
IT WAS REPORTED THAT INTRAOPERATIVELY, WHEN THE SURGEON WAS PLACING THE SCREW ON THE PLATE, THE HEAD OF THE SCREW WAS BROKEN, LEAVING ONE PART ENCRUSTED AND THE OTHER OUTSIDE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8062806 ↗
Other Plate, fixation, bone devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page