POLARSTEM 75102207Recalls & reported problems
Smith & Nephew, Inc. · POLARSTEM COLLAR CALCAR REAMER SMALL · Class 1
All models of POLARSTEM (63)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00885556535516), received between Aug 15, 2019 and Oct 8, 2024 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Dull, Blunt” (1 report).
1 other recall cover devices cleared under the same FDA submission (K143739); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Dull, Blunt1
- Human-Device Interface Problem1
- Material Fragmentation1
- Positioning Failure1
Patient problems
- Bone Fracture(s)3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 8, 2024Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT, DURING A THR SURGERY, THE POLARSTEM COLLAR CALCAR REAMER SMALL HAD METAL FILINGS/DEBRIS. THE AFFECTED AREA WAS HEAVILY IRRIGATED AND SUCTIONED TO REMOVE ALL DEBRIS. ALL THE BROKEN PIECES WERE REMOVED AND ACCOUNTED. THE PROCEDURE WAS COMPLETED, AFTER A NON-SIGNIFICANT DELAY, WITH THE SAME DEVICE. FOLLOWING THE CASE THE REAMER WAS INSPECTED AND WAS FOUND THERE WAS WEAR IN THE CALCAR REAMER AND ON THE TRUNION INSERT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20397827 ↗InjuryNov 13, 2023Positioning Failure; Positioning Failure
IT WAS REPORTED THAT DURING A THR SURGERY, AFTER BROACHING THE FEMORAL CANAL, IT WAS ASKED TO OPEN THE POLARSTEM CALCAR REAMERS. THE NURSE OPENED THE TRAY AND THE SURGEON ATTEMPTED TO REAM ACCORDING TO TECHNIQUE; HOWEVER, THE POLARSTEM CALCAR REAMER 40MM WAS NOT LOWERING ONTO THE POLARSTEM CALCAR REAMER GUIDE AND WAS SWINGING IN THE AIR. THE SURGEON ASKED IF THERE WAS A PROBLEM WITH THE COLLAR OF THE NECK MODULE OR IF IT WAS THE POLARSTEM REAMERS AND NOT FROM OTHER COMPANIES. IT WAS GOT THE SECOND LOAN TRAY AVAILABLE AND WHEN THE SURGEON REPEATED THE PROCESS, THIS TIME THE POLARSTEM CALCAR REAMER 40MM WAS EXTREMELY AGGRESSIVE, CAUSING HIM TO LOSE CONTROL OF THE DRILL AND CAUSE A SMALL FRACTURE IN THE PROXIMAL FEMUR. THIS PROMPTED HIM TO ASK IF THERE WERE ANY CABLES ON THE SHELF THAT HE COULD USE TO SECURE AND FIX IT AND, FORTUNATELY, SYNTHES CABLES WERE AVAILABLE. THE SURGEON USED A CABLE, AND THE CASE WAS RESUMED BY IMPLANTING THE CEMENTED POLARSTEM STEM AND FINALIZING THE CASE. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH THE SAME DEVICE. A SMALL FRACTURE AROUND THE PROXIMAL FEMUR NEAR THE OSTEOTOMY WAS REPORTED AS A CONSEQUENCE OF THIS PROBLEM. IT WAS RESOLVED
Patient problems: Bone Fracture(s)
MAUDE report 18120803 ↗InjuryApr 19, 2022Dull, Blunt; Adverse Event Without Identified Device or Use Problem; Dull, Blunt; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A THR SURGERY THE POLARSTEM COLLAR CALCAR REAMER SMALL WAS DULL AND FRACTURED PATIENT´S CALCAR. THE PROCEDURE WAS COMPLETED, WITHOUT DELAY.
Patient problems: Bone Fracture(s)
MAUDE report 14140477 ↗InjuryAug 15, 2019Human-Device Interface Problem; Human-Device Interface Problem
IT WAS REPORTED THAT DURING SURGERY, WHILE REAMING THE CALCAR BY USING THE CALCAR REAMER, THE FEMORAL BONE WAS BROKEN.
Patient problems: Bone Fracture(s)
MAUDE report 8899760 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K143739. One submission often covers many models, so check the product and lot details in each record.
- Z-2827-2020 · Jul 22, 2020 · Open, ClassifiedPOLARSTEM COLLAR Reamer Guide, Catalog Number: 75102205, UPC Codes: 00885556535493 - Product Usage: Once the appropriate trial rasp size is in place, the adapter is removed and the
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page