LENS 72204685Recalls & reported problems
Smith & Nephew, Inc. · LENS INTEGRATED SYSTEM NON-WIFI VERSION · Class 2
All models of LENS (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00885556617908), received between Aug 9, 2018 and Jun 2, 2021 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Display or Visual Feedback Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Display or Visual Feedback Problem3
- No Display/Image3
- Image Display Error/Artifact1
- Poor Quality Image1
Patient problems
- No Known Impact Or Consequence To Patient7
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryJun 2, 2021Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING USE IN AN ARTHROSCOPIC KNEE SURGERY, ON THE LENS INTEGRATED SYSTEM APPEARS AS A YELLOW SORT OF HAZE ON SCREEN, PRODUCING A SCREEN IMAGE NOT RECOGNIZABLE AND NOT ALLOWING THE SURGERY TO CONTINUE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11924359 ↗MalfunctionAug 21, 2020No Display/Image; No Display/Image
IT WAS REPORTED THAT DURING A SURGERY THE LENS INTEGRATED SYSTEM NON-WIFI VERSION PICTURE DISAPPEARED. THE DEVICE WAS REBOOTED AND IT WORKED FINE. THERE WAS A SOFTWARE INCOMPATIBILITY BETWEEN THE LENS INTEGRATED SYSTEM AND THE CAMERA HEAD. IT IS UNKNOWN IF THERE WAS AN AVAILABLE BACK UP DEVICE OR A SIGNIFICANT DELAY DURING THE PROCEDURE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10437983 ↗MalfunctionAug 21, 2020No Display/Image; No Display/Image
IT WAS REPORTED THAT DURING A SURGERY THE LENS CAMERA HEAD PICTURE DISAPPEARED. THE DEVICE WAS REBOOTED AND IT WORKED FINE. THERE WAS A SOFTWARE INCOMPATIBILITY BETWEEN THE LENS INTEGRATED SYSTEM AND THE CAMERA HEAD. IT IS UNKNOWN IF THERE WAS AN AVAILABLE BACK UP DEVICE OR A SIGNIFICANT DELAY DURING THE PROCEDURE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10437955 ↗MalfunctionAug 19, 2020No Display/Image; No Display/Image
IT WAS REPORTED THAT DURING A SURGERY THE LENS CAMERA HEAD PICTURE DISAPPEARED. OR STAFF SUSPECTS WAS A SOFTWARE INCOMPATIBILITY BETWEEN THE LENS INTEGRATED SYSTEM AND THE CAMERA HEAD. IT IS UNKNOWN IF THERE WAS AN AVAILABLE BACK UP DEVICE OR A SIGNIFICANT DELAY DURING THE PROCEDURE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10428989 ↗MalfunctionSep 4, 2018Poor Quality Image
IT WAS REPORTED THAT SINCE LENS SYSTEM INSTALLED, YELLOW TINTED IMAGE SHOW UP OCCASIONALLY. REPLACED CCU 4 TIMES (4 CCUS FAILED) AND CAMERA HEAD, LIGHT CABLE, SCOPES ALL HAVE BEEN CHANGED AND CHECK CONDITION IS FULLY GOOD. HOWEVER STILL YELLOW IMAGE SHOWS UP. NO PATIENT INJURY WAS REPORTED. BACK UP WAS USED TO COMPLETE THE PROCEDURE. A DELAY GREATER THAN 30 MINUTES WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7844097 ↗MalfunctionSep 4, 2018Display or Visual Feedback Problem
IT WAS REPORTED THAT SINCE LENS SYSTEM INSTALLED, YELLOW TINTED IMAGE SHOW UP OCCASIONALLY. REPLACED CCU 4 TIMES (4 CCUS FAILED) AND CAMERA HEAD, LIGHT CABLE, SCOPES ALL HAVE BEEN CHANGED AND CHECK CONDITION IS FULLY GOOD. HOWEVER STILL YELLOW IMAGE SHOWS UP. NO PATIENT INJURY WAS REPORTED. BACK UP WAS USED TO COMPLETE THE PROCEDURE. A DELAY GREATER THAN 30 MINUTES WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7844038 ↗MalfunctionSep 4, 2018Display or Visual Feedback Problem
IT WAS REPORTED THAT SINCE LENS SYSTEM INSTALLED, YELLOW TINTED IMAGE SHOW UP OCCASIONALLY. REPLACED CCU 4 TIMES (4 CCUS FAILED) AND CAMERA HEAD, LIGHT CABLE, SCOPES ALL HAVE BEEN CHANGED AND CHECK CONDITION IS FULLY GOOD. HOWEVER STILL YELLOW IMAGE SHOWS UP. NO PATIENT INJURY WAS REPORTED. BACK UP WAS USED TO COMPLETE THE PROCEDURE. A DELAY GREATER THAN 30 MINUTES WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7843957 ↗MalfunctionAug 9, 2018Image Display Error/Artifact
IT WAS REPORTED THAT SINCE LENS SYSTEM INSTALLED, YELLOW TINTED IMAGE SHOW UP OCCASIONALLY. THE DEVICE WAS TEST WITH DIFFERENT DEVICES AND THE ISSUE PERSISTS. A SIGNIFICANT DELAY WAS REPORTED. THE PROCEDURE WAS COMPLETED WITH A BACKUP DEVICE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7768914 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page