Skip to content

EVOS 71175069Recalls & reported problems

Smith & Nephew, Inc. · EVOS SMALL 3.5MM DEPTH GAUGE SHORT · Class 2

Quick summary

This Smith & Nephew EVOS Small depth gauge is a bone fixation instrument used to measure screw depth during surgery. Reports mostly describe the tip or hook snapping off mid-procedure, sometimes unnoticed until the case ended, with the missing piece occasionally not found on X-ray. Across 28 reports there are no deaths and 1 injury; in most cases surgery continued with a backup instrument. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
28
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 00885556635872), received between Nov 1, 2019 and Mar 7, 2026 — classified as 1 injury, 27 malfunctions.

The most frequently reported specific device problem is “Break” (15 reports).

1 other recall cover devices cleared under the same FDA submission (K162078); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break15
  • Fracture9
  • Component Missing2
  • Calibration Problem1
  • Device Markings/Labelling Problem1
  • Imprecision1

Patient problems

  • No Clinical Signs, Symptoms or Conditions22
  • No Consequences Or Impact To Patient2
  • No Information2
  • No Known Impact Or Consequence To Patient2
  • Insufficient Information1
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 7, 2026Fracture

    IT WAS REPORTED THAT, DURING AN ORIF SURGERY, THE TIP OF ONE (1) EVOS SMALL 3.5MM DEPTH GAUGE SHORT BROKE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A SIMILAR S+N BACK-UP PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24541645 ↗
  • MalfunctionDec 16, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING AN ORIF FOREARM SURGERY, THE EVOS SMALL 3.5MM DEPTH GAUGE SHORT TIP BROKE OFF DURING NORMAL USE. NO PIECES FELL INSIDE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23822321 ↗
  • MalfunctionApr 26, 2025Break; Break

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION, AN EVOS SMALL 3.5MM DEPTH GAUGE SHORT BROKE IN SURGERY. IT IS UNKNOWN HOW THE SURGERY WAS COMPLETED OR IF THIS CAUSED A SURGICAL DELAY. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21915665 ↗
  • MalfunctionJul 26, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING ORIF SURGERY, THE TIP OF ONE (1) EVOS SMALL 3.5MM DEPTH GAUGE SHORT BROKE AND WAS UNUSABLE. THE PROCEDURE WAS RESUMED USING A S+N BACKUP DEVICE. IT IS UNKNOWN IF THERE WAS SURGICAL DELAY. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19844614 ↗
  • MalfunctionJul 1, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING ORIF SURGERY, THE HOOK TIP OF ONE (1) EVOS SMALL 3.5MM DEPTH GAUGE SHORT BROKE OFF OR WORE OFF. THE DEVICE CANNOT BE USED TO HOOK ON THE FAR CORTEX BECAUSE OF THE AFOREMENTIONED PROBLEM. THE TIP WAS NOT FOUND IN THE WOUND UNDER X-RAY. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING A S+N BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19643988 ↗
  • MalfunctionJan 18, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING AN INTERNAL FIXATION SURGERY, WHILE PLATING, THE TIP OF ONE (1) EVOS SMALL 3.5MM DEPTH GAUGE SHORT SNAPPED OFF. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18536472 ↗
  • MalfunctionJun 13, 2023Fracture; Fracture

    IT WAS REPORTED, THAT THE TIP OF ONE (1) EVOS SMALL 3.5MM DEPTH GAUGE SHORT BROKE OFF DURING THE PROCEDURE. WHEN TRYING TO GET A MEASUREMENT FOR SCREW PLACEMENT. THE SURGERY STAFF DID NOT NOTICE THE PIECE HAD BROKEN OFF THE INSTRUMENT, UNTIL AFTER THE CASE WAS COMPLETE. THE SURGEON WAS NOT CONCERNED ABOUT THE MISSING PIECE. NO X-RAYS ARE AVAILABLE AT THIS TIME. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17122020 ↗
  • MalfunctionMar 21, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING A PLATE PROCEDURE, THE HOOK AT THE END OF A EVOS SMALL 3.5MM DEPTH GAUGE SHORT BROKE OFF AND A RDCE FRCPS W/ PTS {} BRD WAS NOTICED TO BE BENT. HOWEVER, NO PIECES FELL INTO THE PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A SMITH AND NEPHEW BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16583933 ↗
  • MalfunctionAug 17, 2022Fracture; Fracture

    IT WAS REPORTED THAT, UPON INSPECTION FOR A TRAUMA CASE, THE HOOK AT THE TIP OF A EVOS SMALL 3.5MM DEPTH GAUGE SHORT WAS BROKEN OFF. NO SURGICAL DELAY OCCURRED AS A CONSEQUENCE OF THIS ISSUE AND THE PROCEDURE WAS ABLE TO BE CONCLUDED WITH A BACK UP DEVICE. NO PATIENT WAS YET INVOLVED AT THE MOMENT THIS ISSUE WAS IDENTIFIED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15245526 ↗
  • MalfunctionJul 14, 2022Fracture; Fracture

    IT WAS REPORTED THAT BEFORE SURGERY, THE TIP OF THE EVOS SMALL 3.5MM DEPTH GAUGE SHORT WAS NOTICED TO BE BROKEN OFF. A SMITH AND NEPHEW BACK UP DEVICE WAS AVAILABLE. NO INJURY TO PATIENT OR DELAY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15021553 ↗
  • MalfunctionJun 29, 2022Component Missing; Component Missing

    IT WAS REPORTED THAT DURING AN INTERNAL FIXATIONS SURGERY, THE HOOK TIP OF AN EVOS SMALL 3.5MM DEPTH GAUGE SHORT WAS MISSING. SURGERY WAS FINISHED AFTER A NON-SIGNIFICANT DELAY, WITH A SN BACK-UP DEVICE. NO HARM TO THE PATIENT OR FURTHER COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14859730 ↗
  • MalfunctionApr 12, 2022Break; Break

    IT WAS REPORTED THAT THE TIP OF AN EVOS SMALL 3.5MM DEPTH GAUGE SHORT IS BROKEN. SINCE IT WAS NOTICED DURING A FIELD INSPECTION, NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14079946 ↗
  • MalfunctionDec 27, 2021Break; Break

    IT WAS REPORTED THAT THE EVOS SMALL 3.5MM DEPTH GAUGE SHORT WAS BROKEN IN THE WASHER. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13087802 ↗
  • MalfunctionNov 25, 2021Break; Break

    IT WAS REPORTED THAT, DURING TRAUMA SURGERY, THE EVOS SMALL 3.5MM DEPTH GAUGE SHORT BROKE OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12875258 ↗
  • InjuryNov 18, 2021Imprecision; Imprecision

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, AN EVOS SMALL 3.5MM DEPTH GAUGE SHORT WAS NOTICED TO HAVE THE ORANGE DOT MISSING AND IT WAS INCORRECTLY ASSEMBLED WITH A 2.7MM INSERT. THE SCREW LENGTH INCORRECTLY MEASURED WAS INSERTED INTO PATIENT. SURGERY WAS COMPLETED, AFTER A DELAY OF 30 MINUTES OR LESS, WITH A SN BACK-UP DEVICE. NO HARM TO THE PATIENT OR ANY FURTHER COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12836546 ↗
  • MalfunctionSep 2, 2021Break; Break

    IT WAS REPORTED THAT, DURING INSPECTION FOR AN UNKNOWN SURGERY, THE TIP OF A EVOS SMALL 3.5MM DEPTH GAUGE SHORT BROKE OFF. NO SURGICAL DELAY OCCURRED AS A SMITH AND NEPHEW BACK UP DEVICE WAS AVAILABLE TO PERFORM THE PROCEDURE. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12418398 ↗
  • MalfunctionMar 14, 2021Calibration Problem; Calibration Problem

    IT WAS REPORTED THAT, DURING ORIF ANKLE SURGERY IT WAS NOTICED THAT THE EVOS SMALL 3.5MM DEPTH GAUGE SHORT WAS MEASURING WRONG. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A SMITH AND NEPHEW BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11478819 ↗
  • MalfunctionMar 3, 2021Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT DURING SURGERY, THE EVOS SMALL 3.5MM DEPTH GAUGE SHORT DID MEASURE INCORRECTLY. THE PROCEDURE CONTINUED USING A BACKUP DEVICE FROM SMITH AND NEPHEW, WITHOUT SURGICAL DELAY AND NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11407487 ↗
  • MalfunctionFeb 14, 2021Break; Component Missing; Break; Component Missing

    IT WAS REPORTED THAT DURING INSPECTION THE TIP OF THE EVOS SMALL 3.5MM DEPTH GAUGE SHORT WAS SEEN TO BE MISSING. NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11326124 ↗
  • MalfunctionFeb 5, 2021Break; Break

    IT WAS REPORTED THAT, THE TIP OF THE EVOS SMALL 3.5MM DEPTH GAUGE SHORT BROKE OFF. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11284648 ↗
  • MalfunctionDec 7, 2020Break; Break

    IT WAS REPORTED THAT DURING SET UP/ INSPECTION THE TIP OF THE EVOS SMALL 3.5MM DEPTH GAUGE SHORT WAS FOUND TO BE BROKEN. NO PIECES FELL INTO PATIENT. ISSUE WAS NOTICED BEFORE THE CASE SO TIP PROBABLY BROKE OFF DURING STERILE PROCESSING. THERE WAS NO DELAY. A S&N BACK UP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10957040 ↗
  • MalfunctionOct 22, 2020Break; Break

    IT WAS REPORTED THAT THE TIP OF THE EVOS SMALL 3.5 MM DEPTH GAUGE SHORT IS BROKEN. IT IS UNKNOWN WHEN THIS MALFUNCTION OCCURRED AND IF THERE WAS ANY PATIENT INVOLVEMENT. NO INJURIES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10720474 ↗
  • MalfunctionSep 20, 2020Break; Break

    IT WAS REPORTED THAT THE HOOK ON DEPTH GAUGE BROKE OFF. IT IS UNKNOWN WHERE THE EVENT OCCURS.

    Patient problems: No Information

    MAUDE report 10558293 ↗
  • MalfunctionSep 15, 2020Break; Break

    IT WAS REPORTED THAT THE HOOK OF THE DEPTH GAUGE BROKE OFF. UNKNOWN IF S&N BACKUP AVAILABLE, UNKNOWN PROCEDURE. UNKNOWN IF PATIENT INVOLVED.

    Patient problems: Insufficient Information; No Information

    MAUDE report 10538771 ↗
  • MalfunctionAug 6, 2020Break; Break

    IT WAS REPORTED THAT THE DEVICE BROKE DURING SURGERY. USED AN IDENTICAL BACKUP DEVICE TO COMPLETE SURGERY. SURGERY TIME WAS NOT EXTENDED MORE THAN 30 MIN. THE SURGEON DOES NOT FAULT THE DEVICE. THE PATIENT OR SURGICAL PROCEDURE WAS NOT AFFECTED IN ANY WAY DUE TO THE PROBLEM. NO BROKEN PIECES FELL INTO THE PATIENT¿S WOUND AT ALL AND THE DEVICE DID NOT BREAK OVER THE INCISION OR WOUND.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10375887 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K162078. One submission often covers many models, so check the product and lot details in each record.

  • Z-0265-2022 · Oct 25, 2021 · Open, Classified
    Smith and Nephew EVOS Self-Tapping Locking Screw SST: (1)REF 72413575, 3.5MM x 75MM, and (2) REF 72413570, 3.5MM X 70MM. Sterile.

Other Screw, fixation, bone devices

All Screw, fixation, bone devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.