Skip to content

EVOS 72453106Recalls & reported problems

Smith & Nephew, Inc. · EVOS 3.5MM PARTIAL ARTICULAR LATERAL PROXIMAL TIBIA PLATE 6 HOLE LEFT 91MM · Class 2 · Implant

All models of EVOS (190)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885556640319), received between Jan 24, 2025 and Jan 24, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Positioning Failure” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 24, 2025Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION, THE 3.5 LOCKING SCREW WAS INSERTED INTO A LOCKING SCREW HOLE AND PASSED ALL THE WAY THROUGH THE EVOS 3.5MM P/A L-P TIBIA PL 6H L 91MM. THE HOLE IN QUESTION HAD PREVIOUSLY BEEN USED TO COMPRESS THE PLATE TO BONE WITH A CORTICAL SCREW. THE SCREW WAS THEN TRAPPED BENEATH THE PLATE AND NOT REMOVED, IT WAS LEFT IN THE PATIENT. THE PROCEDURE CONTINUED ON WITHOUT FURTHER DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21228885 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page