NA 72204924Recalls & reported problems
Smith & Nephew, Inc. · UNIVERSAL LIGHT GUIDE ORANGE 5MM X 10 FT · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00885556643730), received between Jul 9, 2019 and Nov 6, 2024 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device4
- Mechanical Problem1
Patient problems
- No Patient Involvement3
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 6, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT, DURING A PROCEDURE, THE UNIVERSAL LIGHT GUIDE ORANGE WAS GETTING HOT. SURGERY WAS COMPLETED WITH A SMITH AND NEPHEW BACK-UP DEVICE INSTEAD. THERE WAS DELAY OF 2 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20625916 ↗MalfunctionDec 4, 2019Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT, DURING A PROCEDURE, THE UNIVERSAL LIGHT GUIDE'S CABLE BROKE. THERE WAS NO BACKUP DEVICE AVAILABLE. NO DELAY OR PATIENT INJURY WAS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9413438 ↗MalfunctionJul 9, 2019Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE UNIVERSAL LIGHT GUIDE CABLE WAS DAMAGED AFTER PASSING THROUGH THE STERILIZATION PROCESS AS THEY WERE FOUND STICKING AND OVERHEATING. THE FIRST STERILIZATION WAS MADE IN THE STERRAD; THE SECOND ONE, MADE BY THE HOSPITAL, IS UNKNOWN. AS THIS HAPPENED PRIOR TO ANY CASE, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8774299 ↗MalfunctionJul 9, 2019Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE UNIVERSAL LIGHT GUIDE CABLE WAS DAMAGED AFTER PASSING THROUGH THE STERILIZATION PROCESS AS THEY WERE FOUND STICKING AND OVERHEATING. THE FIRST STERILIZATION WAS MADE IN THE STERRAD; THE SECOND ONE, MADE BY THE HOSPITAL, IS UNKNOWN. AS THIS HAPPENED PRIOR TO ANY CASE, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8774298 ↗MalfunctionJul 9, 2019Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE UNIVERSAL LIGHT GUIDE CABLE WAS DAMAGED AFTER PASSING THROUGH THE STERILIZATION PROCESS AS THEY WERE FOUND STICKING AND OVERHEATING. THE FIRST STERILIZATION WAS MADE IN THE STERRAD; THE SECOND ONE, MADE BY THE HOSPITAL, IS UNKNOWN. AS THIS HAPPENED PRIOR TO ANY CASE, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8774297 ↗
Other Laparoscope, general & plastic surgery devices
- HIGH DEFINITION LCD MONITOR OEV262HOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,937 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-12EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,886 reports
- HD AUTOCLAVABLE CAMERA HEAD OTV-S7ProH-HD-10EOlympus Medical Systems Corp. · Laparoscope, general & plastic surgery1,086 reports
- Auto Suture OMS-T10BTCovidien LP · Laparoscope, general & plastic surgery1 recall955 reports
- PDB OMS-PDB1000Covidien LP · Laparoscope, general & plastic surgery859 reports
- VISERA 4K UHD CAMERA CONTROL UNIT OLYMPUS OTV-S400Olympus Medical Systems Corp. · Laparoscope, general & plastic surgery777 reports
- Auto Suture OMS-T10SBCovidien LP · Laparoscope, general & plastic surgery1 recall697 reports
- Spacemaker SMBTTRNDXCovidien LP · Laparoscope, general & plastic surgery590 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.