Skip to content

EVOS 72468707NRecalls & reported problems

Smith & Nephew, Inc. · EVOS 2.7MM INFERIOR DISTAL CLAVICLE PLATE 7 HOLE LEFT 109MM · Class 2 · Implant

All models of EVOS (190)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885556666142), received between Apr 25, 2023 and Apr 25, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Nonstandard Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 25, 2023Nonstandard Device; Nonstandard Device

    IT WAS REPORTED THAT, DURING FRACTURE FIXATION SURGERY, A PLATE WITH LOCKING HOLES WERE SELECTED TO PROVIDE MORE STABILITY; HOWEVER, THE EVOS 2.7MM INF DIST CLAV PL 7H L 109MM WAS NOT MACHINED PROPERLY AND HAD NO THREADS IN THE HOLES. THIS RESULTED IN THE SCREWS BEING UNABLE TO LOCK INTO THE PLATE. SURGERY WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A SMITH AND NEPHEW BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16808049 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page