Skip to content

JOURNEY 74016424Recalls & reported problems

Smith & Nephew, Inc. · JOURNEY II UNI FEMORALTRIAL HOLDER · Class 1

All models of JOURNEY (566)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00885556692264), received between Apr 25, 2021 and Nov 3, 2025 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Fracture2
  • Detachment of Device or Device Component1
  • Device Fell1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 3, 2025Break; Break

    IT WAS REPORTED THAT, DURING A UKA SURGERY THE PLASTIC OF ONE (1) JRNY II UNI FEMORALTRIAL HOLDER BROKE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A SIMILAR SN BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23459988 ↗
  • MalfunctionOct 28, 2025Break; Break

    IT WAS REPORTED THAT, DURING A UKA SURGERY, THE PLASTIC OF ONE (1) JRNY II UNI FEMORALTRIAL HOLDER WAS BROKEN. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23407610 ↗
  • MalfunctionJun 30, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING A UKR (NAVIO) SURGERY, THE HANDLE OF ONE (1) JRNY II UNI FEMORALTRIAL HOLDER BROKE UPON THE IMPACTING OF THE FEMORAL COMPONENT. ALL PIECES WERE RETAINED OR REMOVED FROM THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22356865 ↗
  • MalfunctionOct 9, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING SURGERY, THE JRNY II UNI FEMORALTRIAL HOLDER HANDLE BROKE/FRACTURED AT THE NECK. IT IS UNKNOWN HOW WAS COMPLETED THE SURGERY AND IF A DELAY WAS PRESENTED. NO INJURY TO PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20407706 ↗
  • InjuryNov 9, 2021Detachment of Device or Device Component; Device Fell; Detachment of Device or Device Component; Device Fell

    IT WAS REPORTED THAT, DURING A TKA SURGERY, THE PIN ON SIDE OF JOURNEY II UNI FEMORALTRIAL HOLDER HOLDING THE SPRING LOADED CLAMPS TO THE SIZE OF THE IMPACTOR FELL OUT. THE HOOK, SPRING AND PIN WERE ALL RECOVERED FROM THE PATIENT WITH TWEEZERS AND FINGERS. SURGERY WAS RESUMED, WITHOUT ANY DELAY, BY CHANGING THE SURGICAL TECHNIQUE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12780026 ↗
  • InjuryApr 25, 2021Break; Break

    IT WAS REPORTED THAT THE KNOB OF THE JRNY II UNI FEMORALTRIAL HOLDER THAT TIGHTENS TO IMPLANT SNAPPED ((B)(4)) AND THE BUMPER CRACKED UPON IMPACTION ((B)(4)). THESE EVENTS OCCURRED DURING USE, WHILE INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED USING THE SAME DEVICES. NO SURGICAL DELAY OR INJURY TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11722110 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates