EVOS 71158103Recalls & reported problems
Smith & Nephew, Inc. · EVOS DISTAL RADIUS VOLAR PLATE 3H LEFT STANDARD TITANIUM 48MM · Class 2 · Implant
All models of EVOS (190)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885556693780), received between Dec 30, 2019 and Jun 24, 2022 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Blocked Connection” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Blocked Connection1
- Connection Problem1
- Deformation Due to Compressive Stress1
- Fitting Problem1
- Flaked1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient1
- No Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJun 24, 2022Flaked; Deformation Due to Compressive Stress; Flaked; Deformation Due to Compressive Stress
IT WAS REPORTED THAT, DURING AN IM NAIL PROCEDURE, WHILE TIGHTENING THE SCREW IN THE UEVOS VOLAR PLATE 3H LEFT STD TI 48MM, THE HOLES OF THE PLATE GOT DAMAGED AND THE ENTIRE SCREW PASSED THROUGH THE PLATE ALONG WITH SOME FRAGMENTS OF METAL. THE SURGERY COULDN'T BE COMPLETED. THE PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14808855 ↗MalfunctionMay 29, 2020Fitting Problem; Blocked Connection; Fitting Problem; Blocked Connection
IT WAS REPORTED THAT DURING SURGERY THE SCREWS WOULDN'T LOCK INTO PLATE. OPENED A NEW SET AND USED THE CONTENTS OF THAT S&N SET. NO DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Information
MAUDE report 10101437 ↗MalfunctionDec 30, 2019Connection Problem; Connection Problem
IT WAS REPORTED THAT DURING SURGERY AN EVOS WRIST TI 4H STD L VOLAR DISTAL RADIUS PLATE (71158104) WAS USED WITH A 16MM LOCKING SCREW (74412416), BUT THE SCREW FELL THROUGH THE PLATE. THEN, AN EVOS WRIST TI 3H STD L VOLAR DISTAL RADIUS (PLATE 71158103) WITH A NEW 16MM SCREW (74412416) WAS TRIED AND IT HAD THE SAME ISSUE. THE SURGERY WAS COMPLETED USING AN OLD-STYLE VOLAR DISTAL RADIUS SET. ADDITIONALLY, 11MM CORTEX SCREW (74402411) WAS FOUND TO HAVE THE SAME FAILURE AS WELL. FOR THIS EVENT, THERE WAS A DELAY GREATER THAN 30 MIN AND THE PATIENT WAS NOT INJURED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9532181 ↗
Other Plate, fixation, bone devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page