EVOS 71158104Recalls & reported problems

Smith & Nephew, Inc. · EVOS DISTAL RADIUS VOLAR PLATE 4H LEFT STANDARD TITANIUM 56MM · Class 2 · Implant

All models of EVOS (190)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00885556693827), received between Dec 30, 2019 and Nov 30, 2023 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Connection Problem1
  • Difficult or Delayed Positioning1
  • Material Deformation1
  • Positioning Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1
  • Rupture1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 30, 2023Material Deformation; Material Deformation

    IT WAS REPORTED THAT, DURING A DISTAL RADIUS OPERATION, TWO DISTAL SCREW HOLES DIDN´T WORK PROPERLY WAY, AS THE LOCKING SCREWS DIDN´T LOCKED INTO TWO OF THE DISTAL HOLES ON THE RIGHT SIDE OF THE EVOS VOLAR PLATE 4H LEFT STD STD TI 56MM. THE EVOS VOLAR PLATE 4H LEFT STD STD TI 56MM WAS INSPECTED AND IT WAS NOTICED THAT THOSE TWO HOLES WERE DAMAGED. SURGERY WAS RESUMED WITH A BACK-UP PLATE, AFTER A SIGNIFICANT DELAY. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18237747 ↗
  • MalfunctionNov 14, 2022Positioning Failure; Positioning Failure

    IT WAS REPORTED THAT, DURING A TRAUMA SURGERY, TWO LOCKING SCREWS WENT THROUGH THE DISTAL HOLES OF THE EVOS VOLAR PLATE 4H LEFT STD, STD TI 56MM DEVICE. THE PROCEDURE WAS RESUMED, AFTER A SIGNIFICANT DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15787668 ↗
  • InjuryApr 21, 2022Difficult or Delayed Positioning; Difficult or Delayed Positioning

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, WHEN INSERTING AN SCREW THROUGH THE PLATE, THE HEAD WENT STRAIGHT THROUGH THE PLATE AND DID NOT ENGAGE WITH THE LOCKING PETALS. SURGERY WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A CHANGE IN THE SURGICAL TECHNIQUE. THE PLATE AND SCREW WAS REMOVED BEFORE AN ADDITIONAL BONE HOLE WAS MADE IN ORDER TO CORRECTLY FIXATE THE SAME PLATE AGAIN. THE SCREW WAS EXPLANTED AND DISCARDED.

    Patient problems: Rupture

    MAUDE report 14165533 ↗
  • MalfunctionDec 30, 2019Connection Problem; Connection Problem

    IT WAS REPORTED THAT DURING SURGERY AN EVOS WRIST TI 4H STD L VOLAR DISTAL RADIUS PLATE (71158104) WAS USED WITH A 16MM LOCKING SCREW (74412416), BUT THE SCREW FELL THROUGH THE PLATE. THEN, AN EVOS WRIST TI 3H STD L VOLAR DISTAL RADIUS (PLATE 71158103) WITH A NEW 16MM SCREW (74412416) WAS TRIED AND IT HAD THE SAME ISSUE. THE SURGERY WAS COMPLETED USING AN OLD-STYLE VOLAR DISTAL RADIUS SET. ADDITIONALLY, 11MM CORTEX SCREW (74402411) WAS FOUND TO HAVE THE SAME FAILURE AS WELL. FOR THIS EVENT, THERE WAS A DELAY GREATER THAN 30 MIN AND THE PATIENT WAS NOT INJURED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9532182 ↗

Other Plate, fixation, bone devices