4KO 72205154Recalls & reported problems
Smith & Nephew, Inc. · 4KO ARTHROSCOPE DV AC 4X30DEGREE X160MM · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00885556736074), received between Apr 6, 2021 and Dec 29, 2025 — classified as 1 injury, 12 malfunctions.
The most frequently reported specific device problem is “Optical Distortion” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Optical Distortion5
- Crack3
- Optical Obstruction3
- Material Integrity Problem1
- Misfocusing1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 29, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT DURING A DEMO PROCEDURE, THE SCOPE LENS WAS DAMAGED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING A SMITH & NEPHEW SIMILAR DEVICE WITH A SURGICAL DELAY GREATER THAN 30 MINUTES. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23908269 ↗InjuryNov 24, 2025Misfocusing; Misfocusing
IT WAS REPORTED THAT, DURING A SHOULDER ARTHROSCOPY PROCEDURE, AN ARTHROSCOPE WAS NOT FOCUSING. THE PROCEDURE WAS RESUMED, WITHOUT ANY SURGICAL DELAY, WITH A CHANGE IN SURGICAL TECHNIQUE, THE DOCTOR DECIDED TO PROCEED WITH THE SURGERY WITHOUT THE ARTHROSCOPY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23630745 ↗MalfunctionSep 9, 2025Optical Obstruction; Optical Obstruction
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, THE ARTHROSCOPE WAS GIVING A FOGGY IMAGE. SURGERY WAS COMPLETED WITH A SMITH AND NEPHEW BACK-UP DEVICE INSTEAD. THERE WAS A SURGICAL DELAY GREATER THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23007355 ↗MalfunctionSep 9, 2025Optical Distortion; Optical Distortion
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY PROCEDURE, THE ARTHROSCOPE DISPLAYED A FOGGY IMAGE. SURGERY WAS COMPLETED WITH A BACK-UP DEVICE INSTEAD. THERE WAS A SURGICAL DELAY GREATER THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23006620 ↗MalfunctionMay 19, 2025Optical Distortion; Optical Distortion
IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, THE ARTHROSCOPE WAS BLURRED. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED AND THERE WAS A SURGICAL DELAY GREATER THAN 30 MINUTES. NO FURTHER COMPLICATION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22047956 ↗MalfunctionMay 14, 2025Optical Obstruction; Optical Obstruction
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE ARTHROSCOPE'S IMAGE WAS FOGGY. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE. THERE WAS A DELAY GREATER THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22019078 ↗MalfunctionApr 1, 2025Optical Obstruction; Optical Obstruction
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, THE SCOPE HAD A FOGGING ISSUE. SURGERY WAS COMPLETED WITH A BACK-UP DEVICE INSTEAD. THERE WAS A SURGICAL DELAY GREATER THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21748943 ↗MalfunctionMar 6, 2025Optical Distortion; Optical Distortion
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, SUDDENLY THE SCOPE WAS BLURRY ON THE SIDE. SURGERY WAS COMPLETED WITH THE SAME DEVICE. THERE WAS A SURGICAL DELAY GREATER THAN 30 MINUTES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21540244 ↗MalfunctionJan 27, 2025Optical Distortion; Optical Distortion
IT WAS REPORTED THAT, DURING SURGERY, THE ARTHROSCOPE DISPLAYED A SHADOW IMAGE WITH BLURRED FOCUS. SURGERY WAS COMPLETED WITH THE SAME DEVICE. THERE WAS A DELAY GREATER THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21248802 ↗MalfunctionDec 30, 2024Optical Distortion; Optical Distortion
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE ARTHROSCOPE VISION WAS NOT CLEAR. THE PROCEDURE WAS COMPLETED USING A SMITH AND NEPHEW BACK UP DEVICE. THERE WAS A SURGICAL DELAY GREATER THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21039615 ↗MalfunctionSep 7, 2023Crack; Crack
IT WAS REPORTED THAT DURING SET UP FOR A MENISCAL REPAIR SURGERY, THE LENS OF THE ARTHROSCOPE WAS CRACKED. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WAS A SIGNIFICANT SURGICAL DELAY GREATER THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17703653 ↗MalfunctionMay 18, 2021Crack; Crack
IT WAS REPORTED THAT DURING SET UP OR INSPECTION, THE DISTAL TIP OF THE SCOPE WAS CRACKED. THERE WAS A S+N BACK-UP DEVICE. NO DELAY WAS REPORTED, AND NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11846484 ↗MalfunctionApr 6, 2021Crack; Crack
IT WAS REPORTED THAT, BEFORE A SHOULDER ARTHROSCOPY, THE SCOPE WAS SCRATCHED. NO DELAY AND A BACK UP WAS AVAILABLE TO COMPLETE THE PROCEDURE. NO OTHER COMPLICATIONS WERE REPORTED. RESULTS OF INVESTIGATION HAVE CONCLUDED THAT THIS UNIT HAD A CRACKED DISTAL LENS WHICH MAKES IT A REPORTABLE EVENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11618808 ↗
Other Arthroscope devices
- FMS 283512Medos International Sàrl · ARTHROSCOPE1,800 reports
- Arthrex® AR-6480Arthrex, Inc. · ARTHROSCOPE949 reports
- FMS VUE 284002Depuy Mitek, LLC · ARTHROSCOPE829 reports
- MITEK 242018Medos International Sàrl · Arthroscope731 reports
- METRx® System 9560100Medtronic Sofamor Danek, Inc. · Arthroscope667 reports
- FMS VUE II 284004Medos International Sàrl · Arthroscope562 reports
- VAPR COOLPULSE 228146Depuy Mitek, LLC · ARTHROSCOPE474 reports
- FMS DUO 284580Depuy Mitek, LLC · ARTHROSCOPE458 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.