EVOS 72582112Recalls & reported problems

Smith & Nephew, Inc. · EVOS 3.5MM/4.5MM PERIPROSTHETIC PROXIMAL FEMUR PLATE L 12H 257MM · Class 2 · Implant

All models of EVOS (190)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885556762240), received between Jan 20, 2025 and Jan 20, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Nonstandard Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Nonstandard Device1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 20, 2025Nonstandard Device; Nonstandard Device

    IT WAS REPORTED THAT, AFTER AN INTERNAL FIXATION PROCEDURE, A FAILURE WAS DETECTED IN THE PATIENT'S SURGERY, SO THE PATIENT HAD TO UNDERGO ANOTHER OPERATION AND WHEN REMOVING AN EVOS 3.5/4.5 PP PROX FEM PL L 12H 257MM, THE CAUSE OF THE PROBLEM WAS REVEALED, A FAILURE IN THE VARIABLE ANGLE HOLES OF THE PLATE WAS FOUND, DUE TO WHICH THE SCREWS HAD NO SUPPORT AND COULD COMPLETELY GO THROUGH THE HOLE. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Failure of Implant

    MAUDE report 21191146 ↗