Skip to content

EVOS 71177733Recalls & reported problems

Smith & Nephew, Inc. · EVOS LARGE TARGETER 3.5MM LINEAR HEX DRIVER SHAFT W/AO QC · Class 1

All models of EVOS (190)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885556767719), received between May 24, 2023 and Jan 2, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 2, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION AN EVOS LARGE TGTR 3.5MM LINEAR HEX DRVR QC BROKE OFF IN A LARGE SCREWDRIVER HANDLE WHILE TIGHTENING A SCREW. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. IT IS UNKNOWN IF PIECES FELL INTO THE PATIENT, BUT NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18428777 ↗
  • MalfunctionJul 20, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING A TRAUMA SURGERY, THE EVOS LARGE TGTR 3.5MM LINEAR HEX DRVR QC DEVICE BROKE OFF INSIDE SCREW HEAD WHILE SURGEON PUT A LOCKING SCREW IN ON POWER. THE TIP OF THE SCREW DRIVER WAS EASILY RETRIEVED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17358224 ↗
  • MalfunctionMay 24, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING SURGERY, THE TIP OF THE EVOS LARGE TGTR 3.5MM LINEAR HEX DRVR QC BROKE OFF IN SCREW HEAD WHILE PUTTING A LOCKING SCREW IN ON POWER. SURGEON DID NOT STOP SOON ENOUGH. THE HEAD WAS EASILY PULLED OUT OF THE SCREW HEAD. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16993301 ↗