Skip to content

EVOS 72585220Recalls & reported problems

Smith & Nephew, Inc. · EVOS 3.5MM/4.5MM PERIPROSTHETIC DISTAL FEMUR SPAN PLATE R 20H 405MM · Class 2 · Implant

All models of EVOS (190)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885556771136), received between Jun 18, 2025 and Jun 18, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration or Expulsion of Device1

Patient problems

  • Deformity/ Disfigurement1
  • Failure of Implant1
  • Loss of fracture reduction1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 18, 2025Migration or Expulsion of Device; Migration or Expulsion of Device

    IT WAS REPORTED THAT, AFTER AN INTERNAL FIXATION SURGERY HAD BEEN PERFORMED ON AN UNKNOWN DATE, THE PLATE AND SCREWS OF ONE (1) EVOS 3.5/4.5 PP DIS FEM PL R 20H 405MM PULLED OUT OF THE PATIENT'S BONE. THIS ADVERSE EVENT WAS SOLVED BY A REVISION SURGERY ON (B)(6) 2025, IN WHICH THE HARDWARE WAS REMOVED. THE CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Deformity/ Disfigurement; Failure of Implant; Loss of fracture reduction; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22288681 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page