Skip to content

REAL INTELLIGENCE ROB10013SRecalls & reported problems

Smith & Nephew, Inc. · REAL INTELLIGENCE ROBOTIC DRILL · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885556780886), received between Sep 28, 2022 and Sep 2, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Connection Problem” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 13, 2023Open, Classified

    Possibility that the user is unable to resolve drill disconnection error messages.

    Z-0708-2024 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Connection Problem2
  • Detachment of Device or Device Component1
  • Difficult to Insert1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 2, 2025Difficult to Insert; Connection Problem; Detachment of Device or Device Component; Difficult to Insert; Connection Problem; Detachment of Device or Device Component

    IT WAS REPORTED THAT, DURING A REAL INTELLIGENCE CORI TKA SURGERY, MANY DIFFICULTIES WERE ENCOUNTERED WHILE ATTACHING THE MILLING CUTTER TO ONE (1) REAL INTELLIGENCE ROBOTIC DRILL. THE SYSTEM REPEATEDLY DISPLAYED AN ERROR MESSAGE DURING THE CALIBRATION OF THE CUTTER, DESPITE THE TRACKERS BEING IN PLACE AND THE POINTER BEING IN CONTACT WITH THE CUTTER. HOWEVER, IT WAS MANAGED TO PROCEED WITH THE PROCEDURE, BUT DURING THE CUTTING PHASE THE CUTTER DETACHED FROM THE RI ROBOTIC DRILL. AFTERWARDS, IT WAS UNABLE TO REATTACH IT. THE PROCEDURE WAS RESUMED USING AN EQUIVALENT S+N BACK UP PRODUCT, AND NO DELAY OR INJURIES WERE REPORTED AS A RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22936170 ↗
  • InjurySep 28, 2022Connection Problem; Connection Problem

    IT WAS REPORTED THAT DURING A CORI ASSISTED TKA PROCEDURE, WHEN THE REAL INTELLIGENCE ROBOTIC DRILL WAS CONNECTED TO THE CONSOLE, CONFIRMATION ICON WAS FLICKERING AND THE DRILL COULD NOT BE RECOGNIZED CORRECTLY. THE CUSTOMER RE-CONNECTED THE DRILL BUT THE FAILURE OCCURRED AGAIN. SURGERY WAS RESUMED AFTER A 30-MINUTE DELAY, WITH A CHANGE IN SURGICAL TECHNIQUE TO NAVIO ASSISTED DUE TO A BACK-UP CORI DRILL WAS NOT AVAILABLE. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15500625 ↗

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.