Skip to content

EVOS 71177760Recalls & reported problems

Smith & Nephew, Inc. · EVOS LARGE 2.5MM DRILL W/AO QC SHORT · Class 1

All models of EVOS (190)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885556791141), received between Mar 25, 2024 and Apr 3, 2026 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3

Patient problems

  • Foreign Body In Patient2
  • Device Embedded In Tissue or Plaque1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 3, 2026Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, ONE (1) EVOS LARGE 2.5MM DRILL W/AO QC SHORT SNAPPED AT THE DISTAL END AT THE POINT WHERE THE FLUTES MEET THE SHAFT. THE INCIDENT OCCURRED WHILE DRILLING THROUGH THE CEMENT MANTLE OF A PROXIMAL FEMUR PERIPROSTHETIC FRACTURE AROUND A HIP REPLACEMENT STEM. THE FLUTED END REMAINED IN THE PATIENT¿S FEMUR, AS THE SURGICAL TEAM DETERMINED THAT REMOVAL WOULD HAVE CAUSED MORE HARM. THE PROCEDURE WAS COMPLETED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient

    MAUDE report 24774252 ↗
  • InjuryApr 29, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING TRAUMA SURGERY ONE (1) EVOS LARGE 2.5MM DRILL W/AO QC SHORT BROKE ON THE FIRST ATTEMPT TO DRILL THE SCREW. THEY WERE NOT REHEATED. THE BROKEN FRAGMENTS OF THE DRILL BITS HAVE REMAINED INSIDE THE PATIENT'S FEMUR. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: Foreign Body In Patient

    MAUDE report 21928067 ↗
  • MalfunctionMar 25, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE EVOS LARGE 2.5MM DRILL W/AO QC SHORT SNAPPED WHILE DRILLING CEMENT USING AND EVOS LARGE PLATE FOR A PERI-PROSTHETIC HIP FRACTURE. THE SURGERY WAS COMPLETED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18969441 ↗

Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices

All ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices →

Sources and updates