Panta 2 Arthrodesis Nail System PAN-1010-0230Recalls & reported problems
Smith & Nephew, Inc. · TARGETING GUIDE · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00885556828878), received between Apr 7, 2025 and Jan 28, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Failure to Align” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Align4
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 28, 2026Failure to Align
IT WAS REPORTED THAT DURING A TTC ARTHRODESIS SURGERY, THE MOST PROXIMAL TIBIAL SCREW HOLE OF THE TARGETING GUIDE WAS SLIGHTLY MISALIGNED. THE DRILL CONTACTED THE NAIL AND PREVENTED THE SCREW FROM BEING PLACED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24205402 ↗MalfunctionAug 26, 2025Failure to Align; Failure to Align
IT WAS REPORTED THAT, DURING A TRAUMA SURGERY, WHILE USING A TARGETING GUIDE, WHENEVER THE NAIL WAS POSITIONED POSTERIORLY IN THE PATIENT, THE PROXIMAL SCREW COULD NOT BE PLACED AS INTENDED. THE MISALIGNMENT OF THE TARGETING GUIDE PREVENTED THE DRILL FROM FITTING PROPERLY, WHICH IN TURN DID NOT ALLOW INSERTION OF THE SCREW. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22895394 ↗MalfunctionAug 26, 2025Failure to Align; Failure to Align
IT WAS REPORTED THAT, DURING A TRAUMA SURGERY, WHILE USING A TARGETING GUIDE, WHENEVER THE NAIL WAS POSITIONED POSTERIORLY IN THE PATIENT, THE PROXIMAL SCREW COULD NOT BE PLACED AS INTENDED. THE MISALIGNMENT OF THE TARGETING GUIDE PREVENTED THE DRILL FROM FITTING PROPERLY, WHICH IN TURN DID NOT ALLOW INSERTION OF THE SCREW. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22894578 ↗MalfunctionAug 22, 2025Failure to Align; Failure to Align
IT WAS REPORTED THAT, DURING A TRAUMA SURGERY, WHILE USING A TARGETING GUIDE, WHEN THE NAIL WAS POSITIONED POSTERIORLY IN THE PATIENT, THE PROXIMAL SCREW COULD NOT BE PLACED AS INTENDED. THE MISALIGNMENT OF THE TARGETING GUIDE PREVENTED THE DRILL FROM FITTING PROPERLY, WHICH IN TURN DID NOT ALLOW INSERTION OF THE SCREW. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22878041 ↗MalfunctionApr 7, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION, AFTER IMPACTION OF THE NAIL, IT WAS DISCOVERED THAT A PIECE OF ONE (1) TARGETING GUIDE BROKE OFF FROM THE TOP. THERE WAS NO IMPACTION TO THE TARGETING GUIDE IN THE AREA IT BROKE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21793029 ↗
Other Orthopedic manual surgical instrument devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page