ENGAGE 1-10013-509Recalls & reported problems

Smith & Nephew, Inc. · TIBIAL INSERT SIZE 5-LEFT MED 9MM · Class 2 · Implant

All models of ENGAGE (80)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885556873670), received between May 8, 2022 and May 26, 2023 — classified as 3 injuries.

The most frequently reported specific device problem is “Malposition of Device” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

4 other recalls cover devices cleared under the same FDA submission (K190439, K230653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. May 2, 2023Class IIOpen, Classified

    Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.

    Z-1673-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Malposition of Device1

Patient problems

  • Pain2
  • Failure of Implant1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, 6 MONTHS AFTER UNICOMPARTMENTAL KNEE ARTHROPLASTY WAS PERFORMED ON AN UNKNOWN DATE, THE PATIENT EXPERIENCED PAIN. THIS ADVERSE EVENT WAS TREATED WITH REVISION SURGERY FROM CEMENTLESS UKA TO CEMENTED UKA ON 21-SEP-2021. DURING THIS SURGERY, THE ANCHOR REMOVAL TOOL TIP BROKE. THE IMPLANTS WERE REMOVED WITHOUT ISSUE AND THERE WAS NO REPORTED DAMAGE TO THE IMPLANTS. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Pain

    MAUDE report 17011713 ↗
  • InjuryMay 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER PARTIAL RIGHT KNEE ARTHROPLASTY WAS PERFORMED ON (B)(6) 2020, THE PATIENT EXPERIENCED UNEXPLAINED PAIN AROUND 5-MONTHS POST-OPERATION. X-RAY AND CT IMAGING SHOWED NO EVIDENCE OF PATIENT INJURY OR DEVICE MALFUNCTION. IMPLANTS WERE INTACT, WELL PLACED, AND THE SURROUNDING BONE HAD NO EVIDENCE OF DAMAGE. CT SHOWED GOOD BONE APPROXIMATION OF THE IMPLANTS. EVEN THOUGH, DUE TO THE PAIN, REVISION SURGERY WAS CONDUCTED ON (B)(6) 2021. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Pain

    MAUDE report 17011329 ↗
  • InjuryMay 8, 2022Malposition of Device; Malposition of Device

    IT WAS REPORTED THAT, AFTER A CEMENTLESS UNICOMPARTMENTAL RIGHT KNEE ARTHROPLASTY HAD BEEN PERFORMED ON (B)(6) 2021, THE PATIENT EXPERIENCED SUBSIDENCE OF THE TIBIAL TRAY AND THE FRACTURE OF THE BLADE REGION OF THE TIBIAL ANCHOR. POSTOPERATIVE X-RAYS CONDUCTED ON AN UNSPECIFIED TIME FRAME SUGGEST A MALPOSITION OF THE FEMORAL COMPONENT RESULTING IN EDGE LOADING OF THE TIBIAL TRAY. THE PATIENT UNDERWENT REVISION SURGERY ON (B)(6) 2022 TO TREAT THIS ADVERSE EVENT. DURING THIS PROCEDURE, ALL THE COMPONENTS FROM THE UKA PROSTHESIS CONSTRUCT WERE EXPLANTED AND REPLACED WITH AN UNSPECIFIED SYSTEM. THE PATIENT'S OUTCOME IS UNKNOWN.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 14332634 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K190439, K230653. One submission often covers many models, so check the product and lot details in each record.

  • Z-2621-2023 · Aug 31, 2023 · Open, Classified
    smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
  • Z-1671-2023 · May 2, 2023 · Open, Classified
    Smith&Nephew ENGAGE Cementless Partial Knee System, Tibial Anchor Stem, Part Numbers: a) REF 1-10011-100 (SIZE 1-2), b) REF 1-10011-200 (SIZE 3-4), c) REF 1-10011-300 (SIZE 5-6
  • Z-1670-2023 · May 2, 2023 · Open, Classified
    Smith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial); b) REF 1-10003-110 (SIZE 1 Left Medial);
  • Z-1672-2023 · May 2, 2023 · Open, Classified
    Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100, SIZE 1-LEFT MEDIAL; b) REF 1-10012-150, SIZE 1-RIGHT MEDIAL; c) REF 1-1

Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates