ENGAGE recallZ-1673-2023 · Smith & Nephew, Inc. · initiated May 2, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smith & Nephew, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.
FDA root cause category: Under Investigation by firm
What the company is doing
Smith & Nephew issued an Urgent Medical Device Recall Notice to its consignees on 05/02/2023 via email and letter. The notice explained the problem and the risk and requested the following: Required Actions: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep, district office, or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customer s account number and name in Section A. 2. Please provide a copy of the attached Physician Communication to the surgical staff. The firm is seeking the return of the device.
Recall details
- Recall number
- Z-1673-2023
- Status
- Open, Classified
- Initiated by the company
- May 2, 2023
- Posted by the FDA
- May 31, 2023
- Recalling firm
- Smith & Nephew, Inc.
- Quantity
- 7531 units
- Product code
- HSX
- FDA submissions
- K190439
- Distribution
- US Nationwide distribution.
Product description
Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Insert, Part Numbers: a) REF 1-10013-109, SIZE 1-LEFT MED, 9MM; b) REF 1-10013-110, SIZE 1-LEFT MED, 10MM; c) REF 1-10013-111, SIZE 1-LEFT MED, 11MM; d) REF 1-10013-113, SIZE 1-LEFT MED, 13MM; e) REF 1-10013-159, SIZE 1-RIGHT MED, 9MM; f) REF 1-10013-160, SIZE 1-RIGHT MED, 10MM; g) REF 1-10013-161, SIZE 1-RIGHT MED, 11MM; h) REF 1-10013-163, SIZE 1-RIGHT MED, 13MM; i) REF 1-10013-209, SIZE 2-LEFT MED, 9MM; j) REF 1-10013-210, SIZE 2-LEFT MED, 10MM; k) REF 1-10013-211, SIZE 2-LEFT MED, 11MM; l) REF 1-10013-213, SIZE 2-LEFT MED, 13MM; m) REF 1-10013-259, SIZE 2-RIGHT MED, 9MM; n) REF 1-10013-260, SIZE 2-RIGHT MED, 10MM; o) REF 1-10013-261, SIZE 2-RIGHT MED, 11MM; p) REF 1-10013-263, SIZE 2-RIGHT MED, 13MM; q) REF 1-10013-309, SIZE 3-LEFT MED, 9MM; r) REF 1-10013-310, SIZE 3-LEFT MED, 10MM; s) REF 1-10013-311, SIZE 3-LEFT MED, 11MM; t) REF 1-10013-313, SIZE 3-LEFT MED, 13MM; u) REF 1-10013-359, SIZE 3-RIGHT MED, 9MM; v) REF 1-10013-360, SIZE 3-RIGHT MED, 10MM; w) REF 1-10013-361, SIZE 3-RIGHT MED, 11MM; x) REF 1-10013-363, SIZE 3-RIGHT MED, 13MM; y) REF 1-10013-409, SIZE 4-LEFT MED, 9MM; z) REF 1-10013-410, SIZE 4-LEFT MED, 10MM; aa) REF 1-10013-411, SIZE 4-LEFT MED, 11MM; bb) REF 1-10013-413, SIZE 4-LEFT MED, 13MM; cc) REF 1-10013-459, SIZE 4-RIGHT MED, 9MM; dd) REF 1-10013-460, SIZE 4-RIGHT MED, 10MM; ee) REF 1-10013-461, SIZE 4-RIGHT MED, 11MM; ff) REF 1-10013-463, SIZE 4-RIGHT MED, 13MM; gg) REF 1-10013-509, SIZE 5-LEFT MED, 9MM; hh) REF 1-10013-510, SIZE 5-LEFT MED, 10MM; ii) REF 1-10013-511, SIZE 5-LEFT MED, 11MM; jj) REF 1-10013-513, SIZE 5-LEFT MED, 13MM; kk) REF 1-10013-559, SIZE 5-RIGHT MED, 9MM; ll) REF 1-10013-560, SIZE 5-RIGHT MED, 10MM; mm) REF 1-10013-561, SIZE 5-RIGHT MED, 11MM; nn) REF 1-10013-563, SIZE 5-RIGHT MED, 13MM; oo) REF 1-10013-609, SIZE 6-LEFT MED, 9MM; pp) REF 1-10013-610, SIZE 6-LEFT MED, 10MM; qq) REF 1-10013-611, SIZE 6-LEFT MED, 11MM; rr) REF 1-10013-613, SIZE 6-LEFT MED, 13MM; ss) REF 1-10013-659, SIZE 6-RIGHT MED, 9MM; tt) REF 1-10013-660, SIZE 6-RIGHT MED, 10MM; uu) REF 1-10013-661, SIZE 6-RIGHT MED, 11MM; vv) REF 1-10013-663, SIZE 6-RIGHT MED, 13MM; Unicondylar knee prothesis
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
a) REF 1-10013-109, UDI/DI 885556873236; b) REF 1-10013-110, UDI/DI 885556873243; c) REF 1-10013-111, UDI/DI 885556873250; d) REF 1-10013-113, UDI/DI 885556873267; e) REF 1-10013-159, UDI/DI 885556873274; f) REF 1-10013-160, UDI/DI 885556873281; g) REF 1-10013-161, UDI/DI 885556873298; h) REF 1-10013-163, UDI/DI 885556873304; i) REF 1-10013-209, UDI/DI 885556873311; j) REF 1-10013-210, UDI/DI 885556873328; k) REF 1-10013-211, UDI/DI 885556873335; l) REF 1-10013-213, UDI/DI 885556873342; m) REF 1-10013-259, UDI/DI 885556873359; n) REF 1-10013-260, UDI/DI 885556873366; o) REF 1-10013-261, UDI/DI 885556873373; p) REF 1-10013-263, UDI/DI 885556873380; q) REF 1-10013-309, UDI/DI 885556873397; r) REF 1-10013-310, UDI/DI 885556873403; s) REF 1-10013-311, UDI/DI 885556873410; t) REF 1-10013-313, UDI/DI 885556873427; u) REF 1-10013-359, UDI/DI 885556873434; v) REF 1-10013-360, UDI/DI 885556873441; w) REF 1-10013-361, UDI/DI 885556873458; x) REF 1-10013-363, UDI/DI 885556873465; y) REF 1-10013-409, UDI/DI 885556873472; z) REF 1-10013-410, UDI/DI 885556873489; aa) REF 1-10013-411, UDI/DI 885556873502; bb) REF 1-10013-413, UDI/DI 885556873519; cc) REF 1-10013-459, UDI/DI 885556873526; dd) REF 1-10013-460, UDI/DI 885556873564; ee) REF 1-10013-461, UDI/DI 885556873601; ff) REF 1-10013-463, UDI/DI 885556873649; gg) REF 1-10013-509, UDI/DI 885556873670; hh) REF 1-10013-510, UDI/DI 885556873717; ii) REF 1-10013-511, UDI/DI 885556873762; jj) REF 1-10013-513, UDI/DI 885556873816; kk) REF 1-10013-559, UDI/DI 885556873854; ll) REF 1-10013-560, UDI/DI 885556873892; mm) REF 1-10013-561, UDI/DI 885556873939; nn) REF 1-10013-563, UDI/DI 885556873984; oo) REF 1-10013-609, UDI/DI 885556874028; pp) REF 1-10013-610, UDI/DI 885556874059; qq) REF 1-10013-611, UDI/DI 885556874097; rr) REF 1-10013-613, UDI/DI 885556874134; ss) REF 1-10013-659, UDI/DI 885556874196; tt) REF 1-10013-660, UDI/DI 885556874233; uu) REF 1-10013-661, UDI/DI 885556874288; vv) REF 1-10013-663, UDI/DI 885556874318; ALL BATCHES
Devices named in this recall
48 devices in the FDA device registry whose identifier appears in the recall's code information.
- ENGAGE 1-10013-509Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall3 reports
- ENGAGE 1-10013-609Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall2 reports
- ENGAGE 1-10013-261Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
- ENGAGE 1-10013-359Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
- ENGAGE 1-10013-410Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall1 report
- ENGAGE 1-10013-559Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall1 report
- ENGAGE 1-10013-560Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
- ENGAGE 1-10013-660Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall1 report
- ENGAGE 1-10013-109Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-110Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- ENGAGE 1-10013-111Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- ENGAGE 1-10013-113Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-159Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-160Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- ENGAGE 1-10013-161Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-163Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-209Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-210Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- ENGAGE 1-10013-211Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- ENGAGE 1-10013-213Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-259Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-260Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- ENGAGE 1-10013-263Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- ENGAGE 1-10013-309Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-310Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-311Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-313Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- ENGAGE 1-10013-360Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- ENGAGE 1-10013-361Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-363Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- ENGAGE 1-10013-409Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- ENGAGE 1-10013-411Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-413Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-459Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-460Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-461Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- ENGAGE 1-10013-463Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-510Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-511Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- ENGAGE 1-10013-513Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-561Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-563Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-610Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-611Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-613Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-659Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall0 reports
- ENGAGE 1-10013-661Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- ENGAGE 1-10013-663Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
Product lines
Other devices under the same FDA submission
36 devices cleared under K190439 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- ENGAGE 1-10003-410Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall10 reports
- ENGAGE 1-10011-300Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall7 reports
- ENGAGE 1-10003-400Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall6 reports
- ENGAGE 1-10012-500Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall6 reports
- ENGAGE 1-10003-510Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall5 reports
- ENGAGE 1-10012-550Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall5 reports
- ENGAGE 1-10003-600Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall4 reports
- ENGAGE 1-10012-450Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall4 reports
- ENGAGE 1-10003-500Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall3 reports
- ENGAGE 1-10003-610Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall3 reports
- ENGAGE 1-10011-200Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall3 reports
- ENGAGE 1-10012-350Smith & Nephew, Inc. · Prosthesis, knee, femorotibial, unicompartmental/unicondylar, uncemented, porous-coated, metal/polymer1 recall3 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 92286).
- Z-1670-2023 · Open, ClassifiedSmith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial); b) REF 1-10003-110 (SIZE 1 Left Medial); …
- Z-1671-2023 · Open, ClassifiedSmith&Nephew ENGAGE Cementless Partial Knee System, Tibial Anchor Stem, Part Numbers: a) REF 1-10011-100 (SIZE 1-2), b) REF 1-10011-200 (SIZE 3-4), c) REF 1-10011-300 (SIZE 5-6…
- Z-1672-2023 · Open, ClassifiedSmith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100, SIZE 1-LEFT MEDIAL; b) REF 1-10012-150, SIZE 1-RIGHT MEDIAL; c) REF 1-1…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page