Skip to content

ENGAGE 1-10013-511Recalls & reported problems

Smith & Nephew, Inc. · ENGAGE TIBIAL INSERT SIZE 5-LEFT MED 11MM · Class 2 · Implant

All models of ENGAGE (80)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

4 other recalls cover devices cleared under the same FDA submission (K190439, K230653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. May 2, 2023Class IIOpen, Classified

    Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.

    Z-1673-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

FDA clearance and classification

Cleared through 510(k) K230653 on May 4, 2023 (decision: Substantially Equivalent). Submitted by Smith & Nephew, Inc. as “Smith & Nephew, Inc. ANTHEM Total Knee System, Genesis Uni Knee System, JOURNEY BCS and II Knee Systems, JOURNEY II CR Knee System, JOURNEY II Uni Knee System, Journey II UK Knee System, JOURNEY II XR Bi-Cruciate Retaining Knee System, JOURNEY PFJ System, JOURNEY Uni Knee System, Legion Hinge Total Knee System, Legion Total Knee System, ZUK Unicompartmental Knee System, ENGAGE Partial Knee System and Porous Patella and Porous Tibial Baseplates”. The FDA review took 56 days. FDA record ↗

Cleared through 510(k) K190439 on Nov 21, 2019 (decision: Substantially Equivalent). Submitted by Engage Uni, LLC as “Engage™ Partial Knee System”. The FDA review took 269 days. FDA record ↗

The FDA regulates product code HSX (prosthesis, knee, femorotibial, non-constrained, cemented, metal/polymer) under 21 CFR 888.3520 as a Class II device. 21 CFR 888.3520 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K190439, K230653. One submission often covers many models, so check the product and lot details in each record.

  • Z-2621-2023 · Aug 31, 2023 · Open, Classified
    smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
  • Z-1671-2023 · May 2, 2023 · Open, Classified
    Smith&Nephew ENGAGE Cementless Partial Knee System, Tibial Anchor Stem, Part Numbers: a) REF 1-10011-100 (SIZE 1-2), b) REF 1-10011-200 (SIZE 3-4), c) REF 1-10011-300 (SIZE 5-6
  • Z-1670-2023 · May 2, 2023 · Open, Classified
    Smith&Nephew ENGAGE Cementless Partial Knee System, Porous Femoral components, Part Numbers: a) REF 1-10003-100 (SIZE 1 Right Medial); b) REF 1-10003-110 (SIZE 1 Left Medial);
  • Z-1672-2023 · May 2, 2023 · Open, Classified
    Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Tray, Part Numbers: a) REF 1-10012-100, SIZE 1-LEFT MEDIAL; b) REF 1-10012-150, SIZE 1-RIGHT MEDIAL; c) REF 1-1

Common questions about the ENGAGE 1-10013-511

Has the ENGAGE 1-10013-511 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1673-2023 (Class II), began on May 2, 2023: Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the ENGAGE 1-10013-511 FDA approved?

It was cleared through 510(k) K190439 on Nov 21, 2019; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the ENGAGE 1-10013-511 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the ENGAGE 1-10013-511, and is it still sold?

The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand ENGAGE, model 1-10013-511). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page