Destino Twist Steerable Guiding Sheath DST0654509Recalls & reported problems
Oscor Inc. · Steerable Guiding Sheath, 6.5F · Class 2
All models of Destino Twist Steerable Guiding Sheath (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00885672004002), received between May 20, 2021 and Jun 12, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K140406); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak2
- Break1
- Detachment of Device or Device Component1
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 12, 2025Fluid/Blood Leak; Fluid/Blood Leak
DURING THE PROCEDURE, THE STROKE (GREEN) ON THE SIDE WAS LEAKING. THE GREEN HUB FROM THE HEMOSTASIS VALVE WAS LEAKING ON THE SIDE. THE PATIENT IS WELL. THE INTERVENTION WAS DONE WITH ANOTHER DESTINO TWIST. THE PROCEDURE WAS AN ENDOLEAK EMBOLIZATION WHICH WAS THE INTERVENTION. THE PATIENT WAS NOT AT RISK. HE IS DOING WELL AND NO FURTHER INTERVENTION WAS NECESSARY. THE TIME DELAY FOR THIS PROBLEM WAS APPROX. 15 - 20 MINUTES. THE CASE WAS REPORTED TO THE SWISS AUTHORITY SWISSMEDICS. DEVICE WILL BE RETURNED TO OSCOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22193654 ↗MalfunctionMar 5, 2024Break; Break
THE CUSTOMER REPORTED THAT WHEN THE SCRUB NURSE RECEIVED THE DESTINO TWIST SHEATH ON HER TABLE, THE LATERAL CANNULA FELL DOWN FROM THE HANDLE OF THE SHEATH. IT SEEMS THE CANNULA WAS NOT CONNECTED TO THE HANDLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18837831 ↗MalfunctionSep 7, 2021Fluid/Blood Leak; Material Fragmentation; Fluid/Blood Leak; Material Fragmentation
IT WAS REPORTED THAT DURING A EMBOLISATION, COIL PROCEDURE THE GREEN HUB DISCONNECTED DURING THE PROCEDURE, DURING THE TRANSITION AT THE HANDLING FROM WHITE TO GRAY, BLOOD PRESSED OUT DURING THE ENTIRE PROCEDURE. THERE WAS NO RISK TO THE PATIENT, THEREFORE, THE PROCEDURE WAS FINISHED WITH ANOTHER DESTINO TWIST.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12434398 ↗MalfunctionMay 20, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A PROCEDURE A GREEN PIECE OF DESTINO TWIST CAME LOOSE. AN IMAGE PROVIDED BY THE CUSTOMER SHOWS THE HEMOSTATIC VALVE DISLODGED FROM THE HANDLE ASSEMBLY. NO PATIENT HARM REPORTED. ULTRASOUND-GUIDED ACCESS OF THE RIGHT COMMON FEMORAL ARTERY WAS PERFORMED USING A 21-GAUGE MICROPUNCTURE SET. THIS WAS TRANSITIONED OVER A WHOLEY WIRE TO A 6 FRENCH SHEATH AND THEN ULTIMATELY TO A 6.5 FRENCH OSCOR TWIST CONFORMABLE SHEATH WHICH WAS ADVANCED INTO THE ABDOMINAL AORTA UNDER FLUOROSCOPIC GUIDANCE. THE PATIENT WAS SYSTEMICALLY HEPARINIZED. THE PATIENT TOLERATED THE PROCEDURE WELL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11855014 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K140406. One submission often covers many models, so check the product and lot details in each record.
- Z-0646-2022 · Dec 22, 2021 · CompletedDestino Twist 14F (also branded as Guidestar 14F)
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page