Skip to content

Destino Twist Steerable Guiding Sheath DST0706509Recalls & reported problems

Oscor Inc. · Steerable Guiding Sheath, 7F · Class 2

All models of Destino Twist Steerable Guiding Sheath (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885672004187), received between Apr 21, 2022 and Apr 21, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 other recall cover devices cleared under the same FDA submission (K140406); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 21, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT ACCORDING TO DR. (B)(6), THE INTRODUCER WAS PUSHED INTO THE AORTA VIA A 0.035 GUIDEWIRE, PROBABLY A LUNDERQUIST. THEN INJECTED VIA A CONTRAST MEDIUM PUMP WITH APPROX.1000 DPI CONTRAST. AT THIS PRESSURE, THE HUB DETACHED FROM THE CASING. THERE WAS A VERY LARGE LOSS OF BLOOD. THE WIRE AND THE SHEATH HAD TO BE REMOVED TOGETHER. THEY FINISHED THE INTERVENTION WITH A NEW SHEATH. THERE WERE NO COMPLICATIONS, BUT THE PROCEDURE WAS DELAYED OVER 30 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14171700 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K140406. One submission often covers many models, so check the product and lot details in each record.

  • Z-0646-2022 · Dec 22, 2021 · Completed
    Destino Twist 14F (also branded as Guidestar 14F)

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page