Destino Twist Steerable Guiding Sheath DST0855517Recalls & reported problems
Oscor Inc. · Steerable Guiding Sheath, 8.5F · Class 2
All models of Destino Twist Steerable Guiding Sheath (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885672006525), received between Dec 6, 2022 and Sep 4, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).
1 other recall cover devices cleared under the same FDA submission (K140406); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Deformation Due to Compressive Stress1
- Mechanical Jam1
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjurySep 4, 2024Mechanical Jam; No Apparent Adverse Event; Mechanical Jam; No Apparent Adverse Event
DURING THE PROCEDURE ENDO LEAK POSITION SEARCH AND RENAL STENT PROLONGATION, THE PHYSICIAN HAD TO USE AN 8.5F DESTINO TWIST. HE INSERTED THE SHEATH AND ONCE ON THE CORRECT POSITION, WANTED TO BEND THE CURVE OF THE DESTINO. AS SOON AS HE DID THIS, THE DEVICE SCRATCHED AND WAS NOT USABLE ANYMORE (MEANING THE CURVE HAS NOT BEEN POSSIBLE TO DO). THE PHYSICIAN DECIDED THEN TO TAKE ANOTHER DEVICE IN 8.5F BUT THE SAME PROBLEM HAPPENED. NO HARM TO THE PATIENT. TWO DEVICES WERE USED FOR THAT PROCEDURE TO BE DONE, INSTEAD OF THAT, NO ADVERSE EVENT. DEVICE TO BE RETURNED TO OSCOR. 07/12/2024: CUSTOMER REPORTED A DELAY IN THE PROCEDURE, ANOTHER SHEATH HAS BEEN TAKEN WHICH BROKE AS WELL. THE PHYSICIAN WENT NERVOUS, AND THEY HAD TO FIGURE OUT ANOTHER SOLUTION TO DO THE PROCEDURE. THE PHYSICIAN ACCESSED THE FEMORAL ARTERY WITH FIRST A SHORT 7F SHEATH AND THEN DIRECTLY THROUGH THE SKIN WENT WITH THE DESTINO. SECOND LOT NUMBER REPORTED FOR DP-21586 WILL BE PROCESSED UNDER (B)(4). SUBMITTED FOR ADMINISTRATIVE PURPOSES: 07/29/2024: RECEIVED CONFIRMATION THAT A DELAY OF MORE THAN 30 MINUTES IN TOTAL IS CONFIRMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20134405 ↗MalfunctionDec 6, 2022Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
IT WAS REPORTED THE DESTINO TWIST WAS BENT PROXIMAL TO THE CURVE AND THE DOCTORS COULD NOT GET THE DEVICES THROUGH THE INTRODUCER (SHEATH); IN THE BEGINNING OF THE PROCEDURE THE DOCTORS HAD NOT NOTICED THE DESTINO TWIST WAS BENT. THEY DECIDED TO SWITCH TO AN APTUS 7F, THIS WORKED, BUT NOT FOR ALL REQUIREMENTS. THEY THEN TRIED USING THE DESTINO TWIST AGAIN, BUT AGAIN, THEY WERE NOT ABLE TO GET THE DEVICES THROUGH THE LUMEN (SHEATH) DUE TO THE KINK BEFORE THE CURVE. THEY DECIDED TO USE A NEW DESTINO TWIST (SAME MODEL AND LOT NUMBER) AND IT WORKED. IT WAS REPORTED THERE WAS MORE THAN A 30 MINUTE DELAY. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15925579 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K140406. One submission often covers many models, so check the product and lot details in each record.
- Z-0646-2022 · Dec 22, 2021 · CompletedDestino Twist 14F (also branded as Guidestar 14F)
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.