Destino Reach Steerable Guiding Sheath DCR1007317Recalls & reported problems
Oscor Inc. · Steerable Guiding Sheath, 10F · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885672007621), received between Jun 27, 2022 and Jun 27, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Mechanical Jam” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 27, 2022Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT A 59 YEAR OLD MALE, WEIGHING 69 KILOGRAMS UNDERWENT A STENT PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) OF THE ARTERIA MESENTERICA SUPERIOR (AMS) PROCEDURE, USING TRANSFEMORAL ACCESS. AFTER TRYING TO INSERT THE STENT (VBX GORE VIABAHN 8-59), THE CURVE OF SHEATH COULD NOT BE DEFLECTED ANYMORE. THE DOCTOR EXCHANGED THE STEERABLE SHEATH, USING ANOTHER DST0707517 MODEL AND SUCCESSFULLY COMPLETED THE PROCEDURE. NO EXTENDED HOSPITALIZATION OR ADDITIONAL SURGICAL INTERVENTION WAS REQUIRED, HOWEVER, THERE WAS A 30 MINUTE DELAY IN THE PROCEDURE. THE PATIENT IS FINE AND NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14828124 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page