Destino Reach Steerable Guiding Sheath DCR1007750Recalls & reported problems
Oscor Inc. · Steerable Guiding Sheath, 10F · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885672007669), received between Dec 31, 2024 and Feb 7, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Difficult to Advance” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Advance1
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Stroke/CVA1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 7, 2025Fluid/Blood Leak; Fluid/Blood Leak
THE CUSTOMER STATED DESPITE BEING UNDER PRESSURIZED BAG OF SALINE, BLOOK WAS BACKFLOWING THROUGH THE FLUSH PORT (SEE ATTACHED PHOTO). BLOOD ASPIRATED AND FLUSH LINE / CATHETER FLUSHED WITH NORMAL SALINE. INCREASING PRESSURE IN BAG APPEARED TO RECTIFY THE ISSUE. THIS IS THE 3RD OCCURRENCE OF BACKFLOW/LEAKAGE REPORTED FOR THAT PARTICULAR LOT AT BASILDON, AND THE 5TH TOTAL INCIDENT. PLEASE ACKNOWLEDGE THIS NEW INCIDENT AND PROVIDE RMA TO RETURN SHEATH FOR EVALUATION. ON 30/JAN/2025: EVENT OCCURRED ON 22/JAN/2025, IN THE UNITED KINGDOM, AND DEVICE LOT NUMBER IS S9130723 (3RD INCIDENT WITH SAME SHEATH LOT NUMBER). ISSUE OCCURRED DURING A VENTRICULAR TACHYCARDIA ABLATION PROCEDURE AND WAS COMPLETED SUCCESSFULLY BY APPLYING MORE PRESSURE TO THE SALINE BAG. THERE WAS NO PATIENT HARM, NO MEDICAL OR SURGICAL INTERVENTION, AND NO PROCEDURE DELAY. PICTURE WAS PROVIDED. DEVICE IS AVAILABLE FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21337210 ↗MalfunctionDec 31, 2024Difficult to Advance; Difficult to Advance
UPDATED EVENT DESCRIPTION 01/28/2025: THIS COMPLAINT IS RELATED TO: OCCLUSION (1); DILATOR WOULD NOT PASS FULLY THROUGH THE ID OF SHEATH (LOT S9298168). THERE WAS NO PATIENT INJURY INVOLVED IN THIS INCIDENT. SHEATH WAS NEVER INSERTED INTO THE PATIENT AS IT FAILED DURING PRE-PROCEDURE SETUP. IT WAS REPLACED WITH A 2ND SHEATH FROM SAME LOT WHICH WORKED FINE. THE SUBJECT SHEATH WAS RETURNED TO INTEGER (B)(4). NOTE: THIS EVENT IS NO LONGER REPORTABLE. IT WAS REPORTED DUE TO THE PREVIOUS EVENT DESCRIPTION STATED THE PATIENT HAD A STROKE. THE NEW EVENT DESCRIPTION FROM OUR CUSTOMER STATES THE PATIENT DID NOT HAVE A STROKE. OUR CUSTOMER INTIALLY REPORTED THE WRONG INFORMATION FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; Stroke/CVA
MAUDE report 21047662 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page