Occlutech® Steerable Guiding Sheath 51GS012Recalls & reported problems
Oscor Inc. · Steerable Guiding Sheath, 12F · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885672009403), received between Oct 27, 2020 and Oct 27, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
- Loss of or Failure to Bond1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 27, 2020Loss of or Failure to Bond; Break; Loss of or Failure to Bond; Break
IT WAS REPORTED THAT DURING THE CLOSURE OF A POSTERO-LATERL MITRAL PARAVALVULAR LEAK PROCEDURE SIDE PORT OF THE FIRST STEERABLE GUIDING SHEATH WAS BROKEN. PHYSICIAN CONTINUED THE PROCEDURE RECONNECTING WITH HIS HANDS THE SIDE PORT TO THE SHEATH. SOON AFTER THAT SHEATH TIP WAS UNABLE TO DEFLECT. PHYSICIAN OPENED SECOND STEERABLE SHEATH BUT AFTER SOME MINUTES INSIDE THE PATIENT SHEATH TIP STOP DEFECTING. PULL WIRE OF THE BOTH SHEATH WAS BROKEN AND STANDING OUT OF THE SHEATH TIP. THERE WAS NO SPECIFIC TECHNIQUE WAS USED FOR INSERTION. PROCEDURE DELAYED FOR 40 MIN AND THERE WAS BLOOD LOSS REPORTED AT THE SIDE PORT DETACHMENT. HOWEVER, THERE WAS NO INTERVENTION REQUIRED TO CONTROL THE BLOOD LOSS. THERE WAS NO PATIENT SIDE EFFECTS WERE REPORTED. NO ADDITIONAL INFORMATION AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10743571 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page