Occlutech Delivery Set III 98DS006Recalls & reported problems

Oscor Inc. · OCCLUTECH DELIVERY SET III, 6F 79CM 45° · Class 2

All models of Occlutech Delivery Set III (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885672010188), received between Jul 26, 2024 and Jul 26, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Difficult to Advance1
  • No Apparent Adverse Event1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 26, 2024Break; Difficult to Advance; No Apparent Adverse Event; Break; Difficult to Advance; No Apparent Adverse Event

    PLEASE BE INFORMED THAT TODAY I FACED AN UNPLEASANT SITUATION AT (B)(6) : ASD PATIENT: 9 MONTHS OLD, 6KG ASD MEASURED 9 MM - DEVICE CHOSEN: ASD10 DUE TO TINY PATIENT DIMENSIONS THE DOWNSIZING OF SHEATH WAS SEEN AS A HUGE ADVANTAGE ASD10 PROCEDURE PACK STATED COMPATIBILITY TO 6F ODS III - SHEATH CHOSEN: ODS III 6F FIRST ODS III 6F SHEATH: HARD TO ADVANCE OCCLUDER & FLEXPUSHER CABLE THROUGH LOADER/ SHEATH ASSEMBLY HUB WITH HEMOSTATIC VALVE & SIDE PORT DETACHED FROM SHEATH EXCHANGED TO SECOND ODS III 6F (SAME LOT#): SAME BEHAVIOR NOT ABLE TO ADVANCE OCCLUDER THROUGH SHEATH/ LOADER ASSEMBLY OCCLUDER - GOT STUCK THIRD SHEATH EXCHANGED TO ODS III 7F + ASD10 EXCHANGED DUE TO HIGH FORCES & UNKNOWN DEVICE INTEGRITY - PROCEDURE WAS SUCCESSFULLY COMPLETED DUE TO THE SHEATH EXCHANGES WITH CAREFUL PRODUCT PREPARATION STEPS ACCORDING TO THE IFU THE PROCEDURE WAS DELAYED BY 40 MIN. THANKS TO MY EXCELLENT RELATIONSHIP TO THE ATTENDING PHYSICIAN DR. (B)(6) AND SENIOR CONSULTANT PROF. (B)(6) I WAS ABLE TO DEESCALATE THIS DELICATE SITUATION IN THE CATH LAB. THIS WOULD HAVE BEEN THE LAST CONVINCING CASE BEFORE THE PEDIATRIC DEPARTMENT WOULD HAVE SWITCHED COMPLETELY TO ODS III DUE TO THE DOWNSIZING AR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19837852 ↗

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates