Occlutech Delivery Set III 98DS008Recalls & reported problems
Oscor Inc. · OCCLUTECH DELIVERY SET III, 8F 79CM 45° · Class 2
All models of Occlutech Delivery Set III (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885672010249), received between Apr 24, 2024 and Sep 27, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Deformation Due to Compressive Stress1
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjurySep 27, 2024Break; No Apparent Adverse Event; Break; No Apparent Adverse Event
WE HAVE A NEW COMPLAINT REGARDING ODSIII ((B)(4)). THE CUSTOMER (B)(6) STATED IN THE INITIAL EMAIL THAT "ODSIII WAS BROKEN (REF: (B)(4); LOT: DP- 19004)". THE PATIENT WAS TREATED FOR AN EMERGENCY IMPLANTATION OF A VSD (VENTRICULAR SEPTAL DEFECT) OCCLUDER AFTER A HEART ATTACK. THE CUSTOMER STATED THAT DURING THE VENTRICULAR SEPTAL DEFECT (VSD) CLOSURE IMPLANTATION PROCEDURE, DIRECTLY WHEN RELEASING THE DEVICE, THE SHEATH CAME APART DURING IMPLANTATION, WHEN PUSHING THROUGH THE OCCLUDER. NO DAMAGE, THE SHEATH COULD BE REPLACED WITH A NEW ONE. THERE IS A DELAY IN THE PROCEDURE OF MORE THAN 40 MINUTES. THERE WAS A MEDICAL INTERVENTION FOR 11. IMPLANTATION OF AN ASD OCCLUDED. FORTUNATELY, THERE WAS NO HARM TO THE PATIENT, THE ENTIRE SYSTEM (SHEATH AND THE LOADED OCCLUDER) COULD BE PULLED OUT AND REPLACED WITH A NEW SHEATH. THE ISSUE RESOLVED BY UPSIZING THE SHEATH. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. SUBMISSION FOR ADMINISTRATIVE PURPOSE AS THERE WAS A DELAY IN THE PROCEDURE OF MORE THAN 40 MINUTES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20322429 ↗MalfunctionApr 24, 2024Break; Break
THE CUSTOMER REPORTED THE SHEATH HAD KINKED. THIS OCCLUTECH DELIVERY SET III (ODSIII) 8F MODEL 98DS008, LOT NUMBER DP-19033 WAS USED IN A VENTRICULAR SEPTAL DEFECT (VSD) CLOSURE PROCEDURE. THE VSD CLOSURE COULD NOT BE FINISHED, AND THE PATIENT MY NEED SURGERY. NO SURGICAL INTERVENTION YET. THE PROCEDURE DELAY WAS NO MORE THAN 30 MINUTES. A SECOND ODSIII WAS USED IN THIS PROCEDURE. DP-19004 IS PROCESSED UNDER (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19176428 ↗MalfunctionApr 24, 2024Break; Deformation Due to Compressive Stress; Break; Deformation Due to Compressive Stress
THE CUSTOMER REPORTED THAT THE SHEATH WAS KINKED, AND THE HEMOSTATIC VALVE WAS TORN OFF. THE PROCEDURE COULD NOT BE COMPLETED, THE FIRST ODSIII WAS KINKED AND PULLED THE OCCLUDER THROUGH THE VENTRICLE SEPTUM DEFECT (VSD) AND HAD TO START FROM THE BEGINNING (LOOP AND SO ONE), SECOND ATTEMPT WITH NEW ODS III: THE OCCLUDER WAS IN PERFECT POSITION, DIRECTLY ON VSD, BUT WE COULD NOT PUSH OR PULL THE FLEX PUSHER BECAUSE THE DISTAL PART OF THE FLEX PUSHER WAS DIRECTLY IN FRONT OF THE KINKED PART. WHEN TRYING TO MOVE THE PUSHER, THE HEMOSTATIC VALVE HAS BEEN TORN OFF! WE DECIDED TO FINISH AND NOT TO TRY THE 3RD TIME, BECAUSE OF THE LENGTH OF THE PROCEDURE. THE VSD CLOSURE COULD NOT BE FINISHED, THE PATIENT MAY NEED SURGERY. NO SURGICAL INTERVENTION YET. ADDITIONAL BRAND NAMES USED; OCCLUTECH DELIVERY SET III, 8 F, 98DS008, LOT: DP-19033 AND DP-19004. SEE IMAGES PROVIDED BY THE CUSTOMER. PROCEDURE DELAY WAS NO MORE THAN 30 MINUTES. DP-19033 WILL BE PROCESSED UNDER COMP-24100. DP-19004 WILL BE PROCESSED UNDER COMP-24101.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19173476 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page