Occlutech Delivery Set III 98DS011Recalls & reported problems

Oscor Inc. · OCCLUTECH DELIVERY SET III, 11F 79CM 45° · Class 2

All models of Occlutech Delivery Set III (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885672010331), received between Jul 12, 2024 and Jul 28, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Air/Gas in Device1
  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Air Embolism1
  • Arrhythmia1
  • Bradycardia1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 28, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50057436 ↗
  • InjuryDec 18, 2025Break; Air/Gas in Device; Break; Air/Gas in Device

    DURING THE ATRIAL SEPTAL DEFECT (ASD) CLOSURE PROCEDURE, THE ASD SIZE WAS MEASURED BY TEE, AND AN ASD OCCLUDER DEVICE SIZE 24 MM AND OCCLUTECH DELIVERY SET III 11F WERE SELECTED. THE EQUIPMENT PREPARED BY THE NURSE WHILE FLUSH SHOWED NO ABNORMALITIES. HOWEVER, WHILE THE PHYSICIAN WAS ABOUT TO DEPLOY THE ASD OCCLUDER DEVICE, THE LOADER DEVICE BROKE INTO TWO SEPARATE PARTS. THIS RESULTED IN AIR ENTERING THROUGH THE FRACTURED PROXIMAL END OF THE LOADER DEVICE INTO THE PATIENT'S CIRCULATION, CAUSING BRADYCARDIA. IMMEDIATELY AFTER THE INCIDENT, THE PHYSICIAN ORDERED A RAPID INFUSION OF 200 ML NSS TO CORRECT THE PATIENT'S ARRHYTHMIA AND PROMPTLY REPLACED BOTH THE DELIVERY SET AND THE LOADER DEVICE. THE PROCEDURE WAS DELAYED FOR APPROXIMATELY 10-15 MINUTES, INCLUDING THE TIME REQUIRED TO OPEN A NEW DEVICE AND TO LOAD NSS TO MANAGE THE BRADYCARDIA. THE PROCEDURE WAS THEN CONTINUED AND SUCCESSFULLY COMPLETED. THE PATIENT IS CURRENTLY IN GOOD CONDITION AND HAS NO ONGOING ISSUES. THE DEVICE WILL BE RETURNED.

    Patient problems: Air Embolism; Arrhythmia; Bradycardia

    MAUDE report 23846455 ↗
  • MalfunctionJul 12, 2024Fluid/Blood Leak; Fluid/Blood Leak

    INITIALLY, THE DOCTOR USED NORMAL SALINE TO FLUSH THE LOADER AND REMOVE THE AIR INSIDE THE TUBE OF THE LOADER. DURING INCOMING INSPECTION AFTER THE DOCTOR LOADED THE ASD INSIDE THE LOADER, THE DOCTOR FOUND AIR BUBBLES CONTINUOUSLY GOING INTO THE TUBE OF THE LOADER. DOCTOR CONTINUOUSLY FLUSHED THE LOADER A FEW TIMES, BUT AIR BUBBLES STILL GOT INTO THE TUBE FROM THE HEMOSTATIC VALVE. THEY OPENED A NEW DELIVERY SYSTEM 12F TO SEE ANY IMPROVEMENT. BUBBLES STILL CAME INTO THE LOADER FROM THE HEMOSTATIC VALVE. THE DOCTOR FINALLY USED A PRESSURE BAG TO CONTINUOUSLY FLUSH NS WATER INTO THE LOADER TO PREVENT AIR BUBBLES GOING INTO THE LOADER TUBE AGAIN AND FINALLY FINISHED THE PROCEDURE. 27/JUN/2024: ADDITIONAL INFORMATION. THIS 98DS011 WAS USED IN AN ATRIAL SEPTAL DEFECT (ASD) PROCEDURE. THERE WAS NO PATIENT DIRECTLY INVOLVED, AS THE EVENT OCCURRED WHEN THE DOCTOR WAS DOING THE DE-AIR PROCEDURE ON THE STERILE TABLE. THE DOCTOR DIDN'T CONFIRM EXACTLY WHAT THE COMPATIBILITY PROBLEM WAS, HOWEVER, THE DOCTOR FEELS THAT THE AIR SHOULD GO EASY INTO THE LOADER AFTER THE PUSHER IS INSERTED INTO THE VALVE. NO DELAY IN PROCEDURE, WAS WITHIN 30 MINUTES. THE ISSUE WAS RESOLVED WHEN THE DOCTOR USED PRES

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19731155 ↗

Other INTRODUCER, CATHETER devices