Occlutech Delivery Set III 98DS012Recalls & reported problems

Oscor Inc. · OCCLUTECH DELIVERY SET III, 12F 79CM 45° · Class 2

All models of Occlutech Delivery Set III (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00885672010355), received between Nov 11, 2023 and Nov 15, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Air/Gas in Device” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Air/Gas in Device4
  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 15, 2024Fluid/Blood Leak; Fluid/Blood Leak

    (B)(6) 2024 DURING DEVICE PREP, FOR CLOSURE OF ATRIAL SEPTAL DEFECT (FASD) IN A PATIENT WITH PULMONARY HYPERTENSION. THE 18MM FENESTRATED ATRIAL SEPTAL DEFECT WAS PULLED BACK IN THE LOADER, AND THE OPERATOR STARTED TO FLUSH TO DE-AIR THE SYSTEM. DESPITE ALL ATTEMPTS, THE LOADER COULD NOT BE DE-AIRED DUE TO AIR ENTERING THE LOADER THROUGH THE PROXIMAL HEMOSTATIC VALVE, WHERE THE PUSHER IS INSERTED. ADDITIONAL ATTEMPTS WERE MADE TO KEEP THE PUSHER STRAIGHT WITH THE LOADER AND THE SALINE BOWL TILTED WITH THE DISTAL END OF THE LOADER SUBMERGED, AS WELL AS HOLDING THE LOADER UPRIGHT WITH THE PUSHER AND LOADER STRAIGHT. THE AIR WAS STILL ENTERING THE LOADER THROUGH THE PROXIMAL HEMOSTATIC VALVE. THE 12F ODSIII SYSTEM WAS ABANDONED, AND AN ODS 12F SYSTEM WAS USED INSTEAD. NO ISSUES OCCURRED WITH THE ODSI SHEATH. THE ISSUE WAS RESOLVED WHEN THE OLDER MODEL ODS L WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A DELAY OF LESS THAN 30 MINUTES. NO MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED. THERE ARE NO PICTURES, OR IMAGES AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20694484 ↗
  • MalfunctionJun 28, 2024Air/Gas in Device; Air/Gas in Device

    INITIALLY, THE DOCTOR USED NORMAL SALINE TO FLUSH THE LOADER AND REMOVE THE AIR INSIDE THE TUBE OF THE LOADER. AFTER THE DOCTOR LOADED THE ASD INSIDE THE LOADER, THE DOCTOR FOUND AIR BUBBLES CONTINUOUSLY GOING INTO THE TUBE OF THE LOADER. DOCTOR CONTINUOUSLY FLUSHED THE LOADER A FEW TIMES, BUT AIR BUBBLES STILL GOT INTO THE TUBE FROM THE HEMOSTATIC VALVE. THEY OPENED A NEW DELIVERY SYSTEM 12F TO SEE ANY IMPROVEMENT. BUBBLES STILL CAME INTO THE LOADER FROM THE HEMOSTATIC VALVE. THE DOCTOR FINALLY USED A PRESSURE BAG TO CONTINUOUSLY FLUSH NS WATER INTO THE LOADER TO PREVENT AIR BUBBLES GOING INTO THE LOADER TUBE AGAIN AND FINALLY FINISHED THE PROCEDURE. PRODUCT WILL BE RETURNED. 06/20/2024 FOLLOW UP INFORMATION RECEIVED: THE DOCTOR FEELS THAT WHEN THE PUSHER INSERTS INTO THE LOADER, THE AIR BUBBLE EASILY GOES INTO THE LOADER. THE AIR-BUBBLE GOES INTO THE LOADER DURING THE PREPARATION ON STERILE TABLE. THE PATIENT WAS NOT INVOLVED AT THE TIME THE BUBBLES WERE FOUND. ONLY DURING THE PREPARATION PROCEDURE. NO DELAY IN THE PROCEDURE, WITHIN 30 MINUTES. DOCTOR USED PRESSURE BAG TO CONTINUOUSLY FLUSH NORMAL SALINE INTO LOADER TO PREVENT THE BUBBLE INTO THE LOADER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19634371 ↗
  • MalfunctionJan 30, 2024Air/Gas in Device; Air/Gas in Device

    DURING IMPLANT PROCEDURE OF AN ATRIAL SEPTUM DEFECT (ASD) CLOSURE, THE ODS III DELIVERY SYSTEM (12 FRENCH) WAS PREPARED VERY WELL AND VERY PROFESSIONALLY. THE PHYSICIAN IS TRAINED TO THE DELIVERY SYSTEM III VERY GOOD, AND HE ALSO HAS USED IT A LOT WITHOUT ANY PROBLEMS. IN THIS CASE HE HAS FLUSHED THE SYSTEM SEVERAL TIMES, BUT HE COULD NOT GET THE LOADER AIR-FREE. WHILE USING THE ODS III, A FIGULLA FLEX II ASD OCCLUDER REF29ASD27, SERIAL NUMBER (B)(6) WAS INSERTED THROUGH THE SHEATH/LOADER AT THE TIME OF THE EVENT. THE PHYSICIAN DECIDED TO SWITCH TO THE ODS I DELIVERY SYSTEM 12F. WHEN USING THE ODS I DELIVERY SYSTEM 12F THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED, HOWEVER THERE WAS A 3- OR 4-MINUTE DELAY, WHEN CHANGING OUT THE ODS III FOR THE ODS I DEVICE. AFTER THE PROCEDURE THE PATIENT ALSO FELT GOOD WITHOUT ANY RESTRICTION AND LEFT THE CATH LAB WITHOUT ANY COMPLAINTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18606381 ↗
  • MalfunctionDec 5, 2023Air/Gas in Device; Air/Gas in Device

    DURING AN ASD PROCEDURE, THERE WAS A PROBLEM, AIR GOT INSIDE THE HEMOSTAIC VALVE; IT TOOK ON AIR DURING PREPARATION OF THE DELIVERY SET. THE DELIVERY SET WAS REPLACED WITH ANOTHER DELIVERY SET, SAME MODEL AND SAME LOT NUMBER. THE OCCLUDE WAS INSERTED THROUGH THE SHEATH DURING THE PROCEDURE AND THE IMPLANTATION WAS SUCCESSFUL. THERE WAS NO PROCEDURE DELAY AND NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18268027 ↗
  • MalfunctionNov 11, 2023Air/Gas in Device; Air/Gas in Device

    THE EVENT OCCURRED PRIOR TO PATIENT CONTACT, DURING DEVICE PREPARATION BUT BEFORE DEVICE POSITIONING OF AN ODS III. WHILE FLUSHING THE LOADER WITH THE DEVICE INSIDE, THERE WAS AIR ENTERING FROM THE HAEMOSTATIC VALVE. IT COULD BE REMOVED, BUT AFTER THAT, THE PHYSICIAN DIDN'T TRUST THE VALVE ANYMORE AND WANTED TO USE AN ODS I. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH ANOTHER DEVICE. THERE WERE NO REPORTED ADVERSE SIDE EFFECTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18117046 ↗

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →