Occlutech Delivery Set III 98DS012Recalls & reported problems
Oscor Inc. · OCCLUTECH DELIVERY SET III, 12F 79CM 45° · Class 2
All models of Occlutech Delivery Set III (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00885672010355), received between Nov 11, 2023 and Nov 15, 2024 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Air/Gas in Device” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Air/Gas in Device4
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 15, 2024Fluid/Blood Leak; Fluid/Blood Leak
(B)(6) 2024 DURING DEVICE PREP, FOR CLOSURE OF ATRIAL SEPTAL DEFECT (FASD) IN A PATIENT WITH PULMONARY HYPERTENSION. THE 18MM FENESTRATED ATRIAL SEPTAL DEFECT WAS PULLED BACK IN THE LOADER, AND THE OPERATOR STARTED TO FLUSH TO DE-AIR THE SYSTEM. DESPITE ALL ATTEMPTS, THE LOADER COULD NOT BE DE-AIRED DUE TO AIR ENTERING THE LOADER THROUGH THE PROXIMAL HEMOSTATIC VALVE, WHERE THE PUSHER IS INSERTED. ADDITIONAL ATTEMPTS WERE MADE TO KEEP THE PUSHER STRAIGHT WITH THE LOADER AND THE SALINE BOWL TILTED WITH THE DISTAL END OF THE LOADER SUBMERGED, AS WELL AS HOLDING THE LOADER UPRIGHT WITH THE PUSHER AND LOADER STRAIGHT. THE AIR WAS STILL ENTERING THE LOADER THROUGH THE PROXIMAL HEMOSTATIC VALVE. THE 12F ODSIII SYSTEM WAS ABANDONED, AND AN ODS 12F SYSTEM WAS USED INSTEAD. NO ISSUES OCCURRED WITH THE ODSI SHEATH. THE ISSUE WAS RESOLVED WHEN THE OLDER MODEL ODS L WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A DELAY OF LESS THAN 30 MINUTES. NO MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED. THERE ARE NO PICTURES, OR IMAGES AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20694484 ↗MalfunctionJun 28, 2024Air/Gas in Device; Air/Gas in Device
INITIALLY, THE DOCTOR USED NORMAL SALINE TO FLUSH THE LOADER AND REMOVE THE AIR INSIDE THE TUBE OF THE LOADER. AFTER THE DOCTOR LOADED THE ASD INSIDE THE LOADER, THE DOCTOR FOUND AIR BUBBLES CONTINUOUSLY GOING INTO THE TUBE OF THE LOADER. DOCTOR CONTINUOUSLY FLUSHED THE LOADER A FEW TIMES, BUT AIR BUBBLES STILL GOT INTO THE TUBE FROM THE HEMOSTATIC VALVE. THEY OPENED A NEW DELIVERY SYSTEM 12F TO SEE ANY IMPROVEMENT. BUBBLES STILL CAME INTO THE LOADER FROM THE HEMOSTATIC VALVE. THE DOCTOR FINALLY USED A PRESSURE BAG TO CONTINUOUSLY FLUSH NS WATER INTO THE LOADER TO PREVENT AIR BUBBLES GOING INTO THE LOADER TUBE AGAIN AND FINALLY FINISHED THE PROCEDURE. PRODUCT WILL BE RETURNED. 06/20/2024 FOLLOW UP INFORMATION RECEIVED: THE DOCTOR FEELS THAT WHEN THE PUSHER INSERTS INTO THE LOADER, THE AIR BUBBLE EASILY GOES INTO THE LOADER. THE AIR-BUBBLE GOES INTO THE LOADER DURING THE PREPARATION ON STERILE TABLE. THE PATIENT WAS NOT INVOLVED AT THE TIME THE BUBBLES WERE FOUND. ONLY DURING THE PREPARATION PROCEDURE. NO DELAY IN THE PROCEDURE, WITHIN 30 MINUTES. DOCTOR USED PRESSURE BAG TO CONTINUOUSLY FLUSH NORMAL SALINE INTO LOADER TO PREVENT THE BUBBLE INTO THE LOADER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19634371 ↗MalfunctionJan 30, 2024Air/Gas in Device; Air/Gas in Device
DURING IMPLANT PROCEDURE OF AN ATRIAL SEPTUM DEFECT (ASD) CLOSURE, THE ODS III DELIVERY SYSTEM (12 FRENCH) WAS PREPARED VERY WELL AND VERY PROFESSIONALLY. THE PHYSICIAN IS TRAINED TO THE DELIVERY SYSTEM III VERY GOOD, AND HE ALSO HAS USED IT A LOT WITHOUT ANY PROBLEMS. IN THIS CASE HE HAS FLUSHED THE SYSTEM SEVERAL TIMES, BUT HE COULD NOT GET THE LOADER AIR-FREE. WHILE USING THE ODS III, A FIGULLA FLEX II ASD OCCLUDER REF29ASD27, SERIAL NUMBER (B)(6) WAS INSERTED THROUGH THE SHEATH/LOADER AT THE TIME OF THE EVENT. THE PHYSICIAN DECIDED TO SWITCH TO THE ODS I DELIVERY SYSTEM 12F. WHEN USING THE ODS I DELIVERY SYSTEM 12F THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO MEDICAL OR SURGICAL INTERVENTION WAS REQUIRED, HOWEVER THERE WAS A 3- OR 4-MINUTE DELAY, WHEN CHANGING OUT THE ODS III FOR THE ODS I DEVICE. AFTER THE PROCEDURE THE PATIENT ALSO FELT GOOD WITHOUT ANY RESTRICTION AND LEFT THE CATH LAB WITHOUT ANY COMPLAINTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18606381 ↗MalfunctionDec 5, 2023Air/Gas in Device; Air/Gas in Device
DURING AN ASD PROCEDURE, THERE WAS A PROBLEM, AIR GOT INSIDE THE HEMOSTAIC VALVE; IT TOOK ON AIR DURING PREPARATION OF THE DELIVERY SET. THE DELIVERY SET WAS REPLACED WITH ANOTHER DELIVERY SET, SAME MODEL AND SAME LOT NUMBER. THE OCCLUDE WAS INSERTED THROUGH THE SHEATH DURING THE PROCEDURE AND THE IMPLANTATION WAS SUCCESSFUL. THERE WAS NO PROCEDURE DELAY AND NO PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18268027 ↗MalfunctionNov 11, 2023Air/Gas in Device; Air/Gas in Device
THE EVENT OCCURRED PRIOR TO PATIENT CONTACT, DURING DEVICE PREPARATION BUT BEFORE DEVICE POSITIONING OF AN ODS III. WHILE FLUSHING THE LOADER WITH THE DEVICE INSIDE, THERE WAS AIR ENTERING FROM THE HAEMOSTATIC VALVE. IT COULD BE REMOVED, BUT AFTER THAT, THE PHYSICIAN DIDN'T TRUST THE VALVE ANYMORE AND WANTED TO USE AN ODS I. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH ANOTHER DEVICE. THERE WERE NO REPORTED ADVERSE SIDE EFFECTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18117046 ↗
Other INTRODUCER, CATHETER devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page