Baylis Med Tech Passport Steerable Sheath PSS0705509Recalls & reported problems

Oscor Inc. · Passport Steerable Sheath 7F · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885672012557), received between Apr 21, 2026 and Apr 21, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).

1 other recall cover devices cleared under the same FDA submission (K140406); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deformation Due to Compressive Stress1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 21, 2026Deformation Due to Compressive Stress; Mechanical Jam

    THERE IS A 60-MINUTE PROCEDURE DELAY WITH THIS EVENT. DEFLECTION CODE ADDED AFTER THE MEETING WITH CUSTOMER ON 07APR2026. CUSTOMER ORGANIZATION: (B)(6). EVENT ORIGIN COUNTRY: CANADA. IS PRODUCT AVAILABLE TO BE RETURNED FOR INVESTIGATION? CUSTOMER SENT IN RMA1202123965, THE 3 PARTS WITH BLOOD. COMPLAINT NOTIFIED DATE TO BMT: 31-MAR-2026. ISSUE DESCRIPTION: INFORMATION RECEIVED FROM CUSTOMER VIA EMAIL ON 31-MAR-2026. "TODAY WE HAD 3 7FR PASSPORT STEERABLE SHEATHS THAT DID NOT WORK. THEY KINKED WHEN WE TRIED TO FLEX THEM. I THOUGHT IT WAS DUE TO OPERATOR ERROR HOWEVER WHEN YOU FLEXED THE SHEATH AS YOU ARE SUPPOSED TO DO IT FLEXED INTO A TWISTED CONFIGURATION MAKING IT UNUSABLE. INSTEAD OF 1 WE HAD TO USE 4 AND ONLY WHEN WE USED THE 7FR 17 MM ONE DID WE ACTUALLY HAVE ANY USEFUL FUNCTION. I AM NOT SURE IF THIS IS A BAD LOT OF SHEATHS, BUT THIS IS UNACCEPTABLE PERFORMANCE. IT DELAYED THE CASE BY OVER AN HOUR DUE TO THIS PROBLEM. DR. (B)(6) (SURGEON)". UPDATED INFORMATION ON 2-APR-2026: "ON THE SAME DAY 3 OTHER UNITS OF THE SAME LOT NUMBER SEEMED TO BE UNDAMAGED IN PACKAGING, BUT WHEN THEY TRIED TO DEFLECT THEM, THEY WERE NOT DEFLECTING PROPERLY EVER SO SLIGHTLY BENDING IN THE SHAFT OF TH

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24944758 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K140406. One submission often covers many models, so check the product and lot details in each record.

  • Z-0646-2022 · Dec 22, 2021 · Completed
    Destino Twist 14F (also branded as Guidestar 14F)