N/A (REF 60-7070-104)Recalls & reported problems
Stryker Sustainability Solutions, Inc. · Tourniquet Cuff Dual Port, Single Bladder (Green), PLC · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00885825013905), received between Jan 11, 2018 and Nov 6, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 6, 2024Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE CUFF WOULD NOT INFLATE. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20625824 ↗MalfunctionJan 11, 2018Leak/Splash; Leak/Splash
IT WAS REPORTED THE PRESSURE TOURNIQUET CUFF FAILED DURING ORTHOPEDIC ANKLE PROCEDURE AND WOULD NOT SUSTAIN PRESSURE. EXTENDED PROCEDURE TIME WAS REPORTED. THE DURATION OF EXTENDED PROCEDURE TIME WAS NOT REPORTED. MULTIPLE ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION. NO INFORMATION WAS PROVIDED. HOWEVER, THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7184905 ↗
Other TOURNIQUET, PNEUMATIC devices
- N/AStryker Sustainability Solutions, Inc. · TOURNIQUET, PNEUMATIC1 recall82 reports
- Color Cuff 5921018135Stryker Corporation · TOURNIQUET, PNEUMATIC2 recalls31 reports
- N/AStryker Sustainability Solutions, Inc. · TOURNIQUET, PNEUMATIC1 recall72 reports
- Color Cuff 5921018235Stryker Corporation · TOURNIQUET, PNEUMATIC2 recalls20 reports
- N/AStryker Sustainability Solutions, Inc. · TOURNIQUET, PNEUMATIC1 recall38 reports
- SmartPump 5920011000Stryker Corporation · TOURNIQUET, PNEUMATIC72 reports
- A.T.S.® Personalized Pressure® 60320010100Zimmer Orthopaedic Surgical Products, Inc. · TOURNIQUET, PNEUMATIC1 recall20 reports
- Color Cuff 5921-034-135NSStryker Corporation · TOURNIQUET, PNEUMATIC1 recall20 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page