N/A (REF B5LT)Recalls & reported problems
Stryker Sustainability Solutions, Inc. · Bladeless Trocar & Sleeve · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00885825014193), received between Oct 31, 2022 and May 8, 2026 — classified as 3 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Fracture” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 19, 2026Class IIOpen, Classified
Reusable access port system may crack or seperate during use.
Z-3186-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Fracture5
- Break2
- Material Integrity Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Foreign Body In Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 8, 2026Fracture
IT WAS REPORTED A TROCAR ARRIVED BROKEN IN PACKAGING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS 5 MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25125273 ↗MalfunctionMay 6, 2026Break
IT WAS REPORTED THE TROCAR WAS BROKEN UPON OPENING, FELL APART. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25093154 ↗MalfunctionMay 1, 2026Fracture
IT WAS REPORTED, "BROKEN TROCAR." IT WAS FURTHER REPORTED THE DEVICE WAS BROKEN WHEN THEY OPENED THE PACKAGING AND WAS NOT USED IN A PROCEDURE. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25056535 ↗MalfunctionApr 17, 2026Fracture
IT WAS REPORTED THE TROCAR FELL APART INSIDE THE PATIENT. THEY WERE ABLE TO RETRIEVE THE PIECES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION, AND THE DURATION OF THE EXTENDED PROCEDURE TIME WAS NOT KNOWN BUT THEY HAD TO GET THE PIECES OUT OF THE PATIENT. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24914895 ↗InjuryDec 19, 2025Fracture; Fracture
IT WAS REPORTED THAT THE 5MM PLASTIC TROCAR TIP BROKE OFF. AN X-RAY WAS TAKEN WITH NO EVIDENCE OF PRESENCE IN THE ABDOMEN. THE PATIENT WENT FROM PACU TO CT. AS REPORTED, THE PATIENT HAD TO HAVE A CT SCAN POST-OP. THERE WAS NO OTHER PATIENT INJURY OR MEDICAL INTERVENTION REPORTED AND THERE WAS NO EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23850663 ↗MalfunctionMar 10, 2025Fracture; Fracture
IT WAS REPORTED THE TIP OF THE TROCAR BROKE OFF. AS REPORTED, THE FRACTURED PIECE BROKE OFF BEFORE USE ON THE PATIENT. THE TIP OF THE DEVICE WAS REPORTED TO NOT HAVE BROKEN OFF COMPLETELY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21568463 ↗InjuryApr 22, 2024Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THE CLEAR TIP AT THE END OF THE COMPLAINT DEVICE BROKE OFF INTO THE PATIENT AND, ALTHOUGH THE TEAM DID THEIR DID DILIGENCE IN THE OR TO LOOK FOR IT, IT WAS NOT IDENTIFIED VIA ULTRASOUND UNTIL AFTER THE PATIENT LEFT THE OR. IT WAS ADDITIONALLY REPORTED THAT THE PATIENT WAS BROUGHT BACK INTO THE OR A FEW DAYS LATER TO REMOVE THE PIECE DURING A SECOND SURGERY. THERE WAS NO OTHER PATIENT INJURY OR MEDICAL INTERVENTION REPORTED AND THE COMPLAINANT IS NOT AWARE OF THE EXTENDED PROCEDURE DURATION. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19160764 ↗InjuryOct 31, 2022Break; Break
IT WAS REPORTED THAT THE "TIP OF OBTURATOR TROCAR BROKE OFF IN THE PATIENT DURING THE "OPTIVIEW" PORTION OF THE PROCEDURE." THE TIP GOT STUCK IN THE PATIENT'S FASCIA. THE DOCTOR USED AN X-RAY, AND THE CASE TURNED INTO AN OPEN PROCEDURE TO RETRIEVE THE BROKEN PIECES. IT WAS REPORTED THAT THEY COULD NOT FIND THE BROKEN PIECES. EXTENDED PROCEDURE TIME WAS REPORTED, BUT NO INFORMATION WAS PROVIDED REGARDING THE LENGTH OF THE EXTENDED PROCEDURE TIME. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: Foreign Body In Patient
MAUDE report 15705157 ↗
FDA clearance and classification
The FDA regulates product code NLM (laparoscope, general & plastic surgery, reprocessed) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the N/A
Has the N/A been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-3186-2026 (Class II), began on Aug 19, 2026: Reusable access port system may crack or seperate during use. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the N/A?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Oct 31, 2022 and May 8, 2026. In those reports, the most common specific device problems are “Fracture” (5), “Break” (2) and “Material Integrity Problem” (1); the most common patient problem is “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the N/A?
None of the reports are classified as deaths, 3 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the N/A, and is it still sold?
The FDA's device database lists Stryker Sustainability Solutions, Inc. as the company responsible for it (brand N/A, model N/A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED devices
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED1 recall9 reports
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED1 recall3 reports
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED1 recall2 reports
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED1 recall0 reports
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED1 recall0 reports
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED1 recall0 reports
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED12 reports
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED7 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page