N/A (REF B11LT)Recalls & reported problems
Stryker Sustainability Solutions, Inc. · Bladeless Trocar & Sleeve · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885825014278), received between Nov 9, 2016 and Oct 23, 2025 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 19, 2026Class IIOpen, Classified
Reusable access port system may crack or seperate during use.
Z-3184-2026 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Break1
- Deformation Due to Compressive Stress1
- Detachment Of Device Component1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Foreign Body In Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 23, 2025Deformation Due to Compressive Stress; Deformation Due to Compressive Stress
IT WAS REPORTED THAT THE TROCAR CAP AND ENTIRE TOP PIECE OF TROCAR FELL APART AS SURGEON WAS PULLING GALLBLADDER OUT OF THE ABDOMEN. AS REPORTED, THE PIECES DID NOT FALL INTO THE PATIENT. THE OUTSIDE PORTION OF THE TROCAR FELL APART OUTSIDE OF THE PATIENT. ALL PIECES WERE RETRIEVED. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS 3-4 MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23368110 ↗InjuryMay 13, 2021Fracture; Fracture
IT WAS REPORTED THE TROCAR BROKE OFF DURING SURGERY. THERE WAS NO PATIENT INJURY. AN X-RAY WAS TAKEN TO LOCATE ANY PIECES, BUT DID NOT SHOW UP ON THE FILM FOR THE RADIOLOGIST TO READ. NO ADDITIONAL INCISIONS WERE MADE OR MADE LARGER WHILE SEARCHING FOR IT. EXTENDED PROCEDURE TIME WAS 15 MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11820630 ↗MalfunctionNov 9, 2016Break; Detachment Of Device Component
IT WAS REPORTED WHEN THE SURGEON WENT TO SEW THE FASCIA, A PIECE OF THE TROCAR WAS DISCOVERED. THE PIECE WAS RETRIEVED AND MATCHED UP TO THE TROCAR TIP TO ENSURE ALL PIECES WERE RECOVERED BEFORE STITCHING UP THE PATIENT. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, AND EXTENDED PROCEDURE TIME REPORTED WAS MINIMAL. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: Foreign Body In Patient
MAUDE report 6089385 ↗
FDA clearance and classification
The FDA regulates product code NLM (laparoscope, general & plastic surgery, reprocessed) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the N/A
Has the N/A been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-3184-2026 (Class II), began on Aug 19, 2026: Reusable access port system may crack or seperate during use. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the N/A?
The FDA's MAUDE database lists 3 adverse event reports for this device, received between Nov 9, 2016 and Oct 23, 2025. In those reports, the most common specific device problems are “Break” (1), “Deformation Due to Compressive Stress” (1) and “Detachment Of Device Component” (1); the most common patient problem is “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the N/A?
None of the reports are classified as deaths, 1 as an injury and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the N/A, and is it still sold?
The FDA's device database lists Stryker Sustainability Solutions, Inc. as the company responsible for it (brand N/A, model N/A). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED devices
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED1 recall9 reports
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED1 recall8 reports
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED1 recall2 reports
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED1 recall0 reports
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED1 recall0 reports
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED1 recall0 reports
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED12 reports
- N/AStryker Sustainability Solutions, Inc. · LAPAROSCOPE, GENERAL & PLASTIC SURGERY, REPROCESSED7 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page