N/A (REF 5921-030-135)Recalls & reported problems
Stryker Sustainability Solutions, Inc. · Tourniquet Color Cuff Single Port, Single Bladder (Royal Blue) Quick Connect · Class 1
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00885825016142), received between Mar 7, 2018 and Feb 20, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Mar 1, 2023Open, Classified
Increased incidence of disposable tourniquet cuffs "not able to achieve desired pressure" and "not holding pressure" due to 1) Cuff fabric tearing during use, and 2) Pinched or unseated O-ring at the connection between the cuff, disposable fill line, and/or cuff hosing, which may lead to loss of vascular occlusion, which may lead to operative site blood loss, hemorrhage, or other complications.
Z-1315-2023 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Leak/Splash3
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Blood Loss1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 20, 2025Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE TOURNIQUET CUFF WOULD NOT INFLATE ON ANY MACHINE. AS REPORTED, THE STAFF ALSO TRIED DIFFERENT INFLATION TUBES BUT IT WOULD NOT WORK. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21423550 ↗MalfunctionOct 14, 2022Leak/Splash; Leak/Splash
IT WAS REPORTED THAT THE TOURNIQUET WAS DEFECTIVE AS IT HAD A HOLE IN THE TUBING. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS 25 MINUTES. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15611159 ↗MalfunctionMar 7, 2018Leak/Splash; Leak/Splash
IT WAS REPORTED THAT A PRESSURE TOURNIQUET CUFF FROM THE OR DID NOT WORK. THE PRESSURE TOURNIQUET CUFF FAILED BEFORE THE SURGERY STARTED. THE TOURNIQUET WOULD NOT HOLD PRESSURE. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION AND EXTENDED PROCEDURE TIME REPORTED WAS 5 MINUTES. THE RISK TO THE PATIENT WAS PREDICTED AS MINIMAL BY THE SURGEON, SO THE DECISION WAS MADE TO CONTINUE WITHOUT THE USE OF A WORKING TOURNIQUET. THERE WAS AN ESTIMATED 20 CC OF BLOOD LOSS AND NO TRANSFUSION WAS REQUIRED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: Blood Loss; No Consequences Or Impact To Patient
MAUDE report 7322429 ↗
Other TOURNIQUET, PNEUMATIC devices
- N/AStryker Sustainability Solutions, Inc. · TOURNIQUET, PNEUMATIC1 recall82 reports
- Color Cuff 5921018135Stryker Corporation · TOURNIQUET, PNEUMATIC2 recalls31 reports
- N/AStryker Sustainability Solutions, Inc. · TOURNIQUET, PNEUMATIC1 recall72 reports
- Color Cuff 5921018235Stryker Corporation · TOURNIQUET, PNEUMATIC2 recalls20 reports
- N/AStryker Sustainability Solutions, Inc. · TOURNIQUET, PNEUMATIC1 recall38 reports
- SmartPump 5920011000Stryker Corporation · TOURNIQUET, PNEUMATIC72 reports
- Color Cuff 5921-034-135NSStryker Corporation · TOURNIQUET, PNEUMATIC1 recall20 reports
- A.T.S.® Personalized Pressure® 60320010100Zimmer Orthopaedic Surgical Products, Inc. · TOURNIQUET, PNEUMATIC1 recall20 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.