Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00886333218516), received between Nov 13, 2020 and Aug 9, 2025 — classified as 2 deaths, 7 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Defective Component2
- Material Twisted/Bent2
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
- Entrapment of Device1
- Failure to Obtain Sample1
- Migration1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Foreign Body In Patient2
- Insufficient Information2
- Exsanguination1
- Hemorrhage/Blood Loss/Bleeding1
- Low Blood Pressure/ Hypotension1
- Skin Tears1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 9, 2025Defective Component; Defective Component
REPORTER STATES "I DO HAVE A DEFECTIVE TLABS KIT WHERE THE PLASTIC SHEATH COVERED OVER THE TOP OF THE NEEDLE AND CONSEQUENTLY THE MD WAS CUT TRYING TO UNSHEATHE THE NEEDLE TIP." ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Skin Tears
MAUDE report 22767517 ↗MalfunctionDec 28, 2023Unraveled Material; Unraveled Material
THE PHYSICIAN WAS PERFORMING A TRANS JUGULAR BIOPSY OF THE LIVER. ON INSERTION OF THE PRODUCT IT DID NOT CUT THE SAMPLE AS REQUIRED AND ON EXAMINATION THE MID SECTION OF THE DEVICE WAS DAMAGED. THE COATING ON THE DEVICE WAS UNRAVELED AS WELL WITH A SHARP PROTRUDING METAL PIECE. AS THE PATIENT HAD A VERY HIGH INR WITH HIGH RISK OF BLEEDING THIS DEVICE WAS REPLACED WITH ANOTHER DEVICE OF THE SAME MODEL WHICH PERFORMED 100% AND ADEQUATE SAMPLES WERE EXTRACTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18411949 ↗DeathNov 16, 2023Material Twisted/Bent; Adverse Event Without Identified Device or Use Problem; Material Twisted/Bent; Adverse Event Without Identified Device or Use Problem
THE REPORTED DEVICE WOULD NOT ACQUIRE ANY TISSUE SAMPLES DURING A TRANSJUGULAR LIVER BIOPSY. WHEN THE DEVICE WAS REMOVED FROM THE PATIENT, THE PROVIDER REPORTED THAT THE DISTAL PORTION OF THE BIOPSY NEEDLE WAS BENT/CURVED AND LOOKED UNUSUAL ALTHOUGH NOTHING ABNORMAL WAS ABSORBED UNDER FLUOROSCOPY DURING THE PROCEDURE. ANOTHER DEVICE WAS USED WITHOUT DIFFICULTY TO OBTAIN THE BIOPSY. FOLLOWING THE BIOPSY PROCEDURE, THE PATIENT HAD INTRAPERITONEAL HEMORRHAGE REQUIRING BLOOD TRANSFUSION AND HEPATIC ARTERY EMBOLIZATION. CT IMAGING DEMONSTRATED A LARGE VOLUME HEMOPERITONEUM AND PHYSICAL EXAM FINDINGS WERE CONCERNING FOR ABDOMINAL COMPARTMENT SYNDROME. THE PATIENT WAS IN THE PROCESS OF BEING TAKEN TO THE OR FOR ABDOMINAL DECOMPRESSION WHERE HE ARRESTED AND SUBSEQUENTLY EXPIRED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Insufficient Information
MAUDE report 18148768 ↗DeathNov 9, 2023Failure to Obtain Sample; Material Twisted/Bent; Failure to Obtain Sample; Material Twisted/Bent
THE REPORTED DEVICE WOULD NOT ACQUIRE ANY TISSUE SAMPLES DURING A TRANSJUGULAR LIVER BIOPSY. WHEN THE DEVICE WAS REMOVED FROM THE PATIENT, THE PROVIDER REPORTED THAT THE DISTAL PORTION OF THE BIOPSY NEEDLE WAS BENT/CURVED AND LOOKED UNUSUAL ALTHOUGH NOTHING ABNORMAL WAS OBSERVED UNDER FLUOROSCOPY DURING THE PROCEDURE. ANOTHER DEVICE WAS USED WITHOUT DIFFICULTY TO OBTAIN THE BIOPSY. FOLLOWING THE BIOPSY PROCEDURE, THE PATIENT HAD INTRAPERITONEAL HEMORRHAGE REQUIRING BLOOD TRANSFUSIONS AND HEPATIC ARTERY EMBOLIZATION. CT IMAGING DEMONSTRATED A LARGE VOLUME HEMOPERITONEUM, AND PHYSICAL EXAM FINDINGS WERE CONCERNING FOR ABDOMINAL COMPARTMENT SYNDROME. THE PATIENT WAS IN THE PROCESS OF BEING TAKEN TO THE OPERATING ROOM FOR ABDOMINAL DECOMPRESSION WHEN HE ARRESTED AND SUBSEQUENTLY EXPIRED.
Patient problems: Exsanguination; Low Blood Pressure/ Hypotension
MAUDE report 18103768 ↗MalfunctionOct 15, 2021Therapeutic or Diagnostic Output Failure
SYSTEM MALFUNCTIONING, DUE TO THE INCORRECT EXPOSURE OF THE TRU-CUT NEEDLE NOTCH. SEE ENCLOSED VIDEO. THE PROCEDURE HAS BEEN COMPLETED, BY DISCARDING THE DEVICE AND USING ANOTHER ONE OF THE SAME CODE/BATCH #. NO PATIENT INJURY.
Patient problems: Insufficient Information
MAUDE report 12643746 ↗MalfunctionJul 8, 2021Break
TIP OF THE PRODUCT BROKE OFF IN THE PATIENT.
Patient problems: Foreign Body In Patient
MAUDE report 12139703 ↗MalfunctionMar 10, 2021Defective Component
LOT# 1451428 ON FEB 9TH THAT THE SHEATH AROUND THE ACTUAL BIOPSY NEEDLE DETACHED FROM ITSELF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11457214 ↗MalfunctionJan 25, 2021Break; Entrapment of Device; Migration
RADIO-OPAQUE BAND ON THE TIP OF THE SHEATH BROKE OFF IN THE PATIENT AND TRAVELED TO A PULMONARY ARTERY IN THE PATIENT. IT'S IN A PERIPHERY ARTERY AND WAS LEFT IN THE PATIENT. PATIENT IS FINE AND HAS NO ISSUES ON TEST.
Patient problems: Foreign Body In Patient
MAUDE report 11227588 ↗MalfunctionNov 13, 2020Detachment of Device or Device Component
TIP (DISTAL 5MM) OF THE SHEATH SEEMED TO TEAR AWAY AND ALMOST DISLODGE DURING THE PROCEDURE. SHEATH STAYED INTACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10840513 ↗
Other Introducer, Catheter devices
- VIZIGO D138502Biosense Webster Inc · INTRODUCER, CATHETER2 recalls4 deaths2,839 reports
- VIZIGO D138501Biosense Webster Inc · INTRODUCER, CATHETER2 recalls2 deaths2,005 reports
- Swartz™ 407451St. Jude Medical, Inc. · INTRODUCER, CATHETER5 deaths728 reports
- Edwards eSheath+ Introducer set 914ESPEdwards Lifesciences LLC · Introducer, Catheter22 deaths532 reports
- Fast-Cath™ 406849St. Jude Medical, Inc. · INTRODUCER, CATHETER420 reports
- VIZIGO D138503Biosense Webster Inc · INTRODUCER, CATHETER362 reports
- Agilis™ G408318St. Jude Medical, Inc. · INTRODUCER, CATHETER333 reports
- Agilis™ NxT Steerable Introducer Dual-Reach™ SI-1371-MSt. Jude Medical, Inc. · Introducer, catheter1 death301 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page