Terumo® Advanced Perfusion System 1 816571Recalls & reported problems
Terumo Cardiovascular Systems Corporation · The large roller pump is a peristaltic pump with a 6 inch (15 cm) raceway that can be used for arterial, cardioplegia, vent, and suction applications. A universl tube clamp mechanism accommodates a variety of tubing sizes, including dual tube sets. · Class 2
All models of Terumo Advanced Perfusion System 1 (31)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Terumo Advanced Perfusion System 1 console is a heart-lung machine used for cardiopulmonary bypass. The 237 reports mostly describe pump output and mechanical problems, such as the pump head stopping without an alarm, clamps sticking and displays not working, and several say an alternate device was used and the procedure was completed. The reports include no injuries and no deaths, and several state there were no adverse consequences for the patient. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 228
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 228 adverse event reports that list this exact device (DI 00886799001370), received between Jun 20, 2018 and Aug 6, 2026 — classified as 228 malfunctions.
The most frequently reported specific device problem is “Output Problem” (84 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Output Problem84
- Mechanical Problem71
- Display or Visual Feedback Problem48
- Power Problem12
- No Display/Image10
- Pumping Stopped10
- Mechanical Jam7
- Noise, Audible5
Patient problems
- No Clinical Signs, Symptoms or Conditions171
- No Consequences Or Impact To Patient40
- Insufficient Information23
- No Patient Involvement11
- No Known Impact Or Consequence To Patient5
Recent adverse event reports
The 25 most recent of 228 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2026Mechanical Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50168025 ↗MalfunctionMar 30, 2026Output Problem
IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE ROLLER PUMP GAVE A 'BELT SLIP' ERROR. THE TEAM USED IT FOR THE REMAINDER OF THE CASE AS IS. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24731309 ↗MalfunctionMar 9, 2026Pumping Stopped
IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE ROLLER PUMP STOPPED WORKING. THE PERFUSIONIST RESTARTED THE ROLLER PUMP AND THE ISSUE RESOLVED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24552253 ↗MalfunctionMar 5, 2026Output Problem
IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE ROLLER PUMP HAD A BELT SLIP ERROR. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24526145 ↗MalfunctionFeb 23, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE ROLLER PUMP DISPLAY WAS BLANK. THE TEAM COULD CONTROL THE ROLLER PUMP FROM THE CENTRAL CONTROL MONITOR (CCM). THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24421834 ↗MalfunctionFeb 19, 2026Output Problem
IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE ROLLER PUMP STOPPED WORKING AND AN ERROR MESSAGE, POSSIBLY STATING 'SERVICE NEEDED', WAS DISPLAYED QUICKLY THEN WENT AWAY. THE PERFUSIONIST RESTARTED THE PUMP AND THE ISSUE RESOLVED. THE TEAM USED THE ROLLER PUMP FOR THE REMAINDER OF THE CASE. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24389807 ↗MalfunctionFeb 5, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE CARDIOPLEGIA ROLLER PUMP DISPLAY WAS BLANK WITH NO BACKLIGHT PRESENT. THE TEAM COULD STILL CONTROL THE PUMP FROM THE LOCAL CONTROLS ON THE PUMP HEAD AND THE CENTRAL CONTROL MONITOR (CCM) TOUCH SCREEN. THE MACHINE WAS RESTARTED, THE PUMP HEAD WAS UNPLUGGED THEN PLUGGED BACK IN, BUT THE ISSUE REMAINED. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24273644 ↗MalfunctionFeb 2, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE ROLLER PUMP WAS EXCHANGED WITH A SUCKER PUMP, BUT THE BUTTONS WERE NOT RESPONDING. THE TEAM WAS ABLE TO TURN THE PUMP ON WITH THE CENTRAL CONTROL MONITOR (CCM) CONTROLS. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24241691 ↗MalfunctionJan 30, 2026Inadequate User Interface
IT WAS REPORTED THAT THE POWER BUTTON ON THE ROLLER PUMP LOCAL DISPLAY WAS NOT WORKING. THE ROLLER PUMP POWER WORKED BUT ONLY WHEN CONTROLLED BY THE CENTRAL CONTROL MONITOR (CCM). THERE WERE NO REPORTED CONSEQUENCES TO THE PATIENT. NO OTHER DETAILS REGARDING THE NATURE OF THIS EVENT WERE PROVIDED.
Patient problems: Insufficient Information
MAUDE report 24223073 ↗MalfunctionJan 30, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE ROLLER PUMP DISPLAY DID NOT TURN ON BUT THE PUMP WAS STILL FUNCTIONAL AND ABLE TO BE CONTROLLED BY THE CENTRAL CONTROL MONITOR (CCM). AFTER MULTIPLE ATTEMPTS TO TURN THE DISPLAY ON, THE TEAM PHYSICALLY AGITATED IT AND THE ISSUE RESOLVED. THERE WAS NO DELAY, AND NO REPORTED CONSEQUENCES TO THE PATIENT. NO OTHER DETAILS REGARDING THE NATURE OF THIS EVENT WERE PROVIDED.
Patient problems: Insufficient Information
MAUDE report 24222861 ↗MalfunctionDec 19, 2025Output Problem; Power Problem; Output Problem; Power Problem
IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR CARDIOPULMONARY BYPASS (CPB), THE ARTERIAL ROLLER PUMP STOPPED TWICE, ONLY DISPLAYING A 'SERVICE PUMP' MESSAGE THE SECOND OCCURRENCE. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS AN UNKNOWN DELAY. THERE WAS NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23857799 ↗MalfunctionDec 18, 2025Output Problem; Output Problem
ADDITIONAL INFORMATION WAS RECEIVED THAT AS A RESULT OF THE ISSUE; AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23844144 ↗MalfunctionDec 16, 2025Mechanical Problem; Output Problem; Mechanical Problem; Output Problem
AFTER USE OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE USER REPORTED THAT CARDIOPLEGIA PUMP DISPLAYED A 'BELT SLIP' MESSAGE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23818280 ↗MalfunctionNov 21, 2025Output Problem; Output Problem
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE ROLLER PUMP DISPLAYED A 'SERVICE PUMP' MESSAGE. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23617553 ↗MalfunctionNov 5, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT DURING PRIMING OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE ARTERIAL ROLLER PUMP HEAD SCREEN WENT BLACK. IT STILL FUNCTIONED OTHERWISE, AND THE FLOW STILL APPEARED AND COULD BE CONTROLLED ON THE CENTRAL CONTROL MONITOR (CCM). AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23480139 ↗MalfunctionOct 21, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB), THE ROLLER PUMP DISPLAY WENT BLANK AND THEN CAME BACK ON WHEN IT WAS TAPPED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23353747 ↗MalfunctionOct 16, 2025Power Problem; Power Problem
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE ROLLER PUMP TEMPORARILY LOST POWER. THE TEAM ADJUSTED/MOVED THE CABLE IN THE BACK AND THE ISSUE RESOLVED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23311154 ↗MalfunctionSep 15, 2025Break; Break
THE SERVICE REPAIR TECHNICIAN (SRT) REPORTED THAT DURING TESTING OF THE DEVICE AT THE SERVICE CENTER, THE ROLLER PUMP LID WAS BROKEN AND THE MAGNET WAS MISSING FROM THE LID. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23057130 ↗MalfunctionSep 2, 2025Display or Visual Feedback Problem; Power Problem; Display or Visual Feedback Problem; Power Problem
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE ROLLER PUMP INTERMITTENTLY WOULD NOT TURN ON WHEN THE SYSTEM WAS POWERED ON. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22939690 ↗MalfunctionAug 25, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE ROLLER PUMP DISPLAY WAS NOT WORKING. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22891535 ↗MalfunctionAug 21, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
THE FIELD SERVICE REPRESENTATIVE (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, SOME OF THE ROLLER PUMP PIXELS WERE NOT ILLUMINATING MAKING IT DIFFICULT TO READ. THE PUMP COULD STILL BE CONTROLLED LOCALLY AND WITH THE CENTRAL CONTROL MONITOR (CCM). THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22864968 ↗MalfunctionAug 18, 2025Power Problem; Power Problem
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE ROLLER PUMP WOULD NOT POWER ON. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22829850 ↗MalfunctionAug 14, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE ROLLER PUMP DISPLAY WAS NOT WORKING. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22805137 ↗MalfunctionAug 6, 2025Output Problem; Output Problem
THE FIELD SERVICE REPRESENTATIVE (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, THE ROLLER PUMP DISPLAYED A 'SERVICE PUMP, MOTOR ERROR' MESSAGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22726888 ↗MalfunctionJul 28, 2025Power Problem; Power Problem
IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE ROLLER PUMP WAS ON BUT WOULD NOT ROTATE. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22632531 ↗
FDA clearance and classification
The FDA regulates product code DTQ (console, heart-lung machine, cardiopulmonary bypass) under 21 CFR 870.4220 as a Class II device. 21 CFR 870.4220 ↗
Devices of this type normally need 510(k) clearance before sale.
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Terumo® Advanced Perfusion System 1 816571
Has the Terumo® Advanced Perfusion System 1 816571 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00886799001370), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Terumo® Advanced Perfusion System 1 816571?
The FDA's MAUDE database lists 228 adverse event reports for this device, received between Jun 20, 2018 and Aug 6, 2026. In those reports, the most common specific device problems are “Output Problem” (84), “Mechanical Problem” (71) and “Display or Visual Feedback Problem” (48). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Terumo® Advanced Perfusion System 1 816571?
None of the reports are classified as deaths, 0 as injuries and 228 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Terumo® Advanced Perfusion System 1 816571, and is it still sold?
The FDA's device database lists Terumo Cardiovascular Systems Corporation as the company responsible for it (brand Terumo® Advanced Perfusion System 1, model 816571). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Console, Heart-Lung Machine, Cardiopulmonary Bypass devices
- Terumo® Advanced Perfusion System 1 801188Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass3 recalls449 reports
- MAQUET CARDIOHELP CARDIOHELP-iMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS5 recalls2 deaths59 reports
- MAQUET CARDIOHELP CARDIOHELP SystemMaquet Cardiopulmonary AG · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS4 recalls16 deaths99 reports
- Terumo® Advanced Perfusion System 1 816300Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass298 reports
- Terumo® Advanced Perfusion System 1 801763Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 death230 reports
- ESSENZ HLMLivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 recall1 death111 reports
- Terumo® Advanced Perfusion System 1 5773Terumo Cardiovascular Systems Corporation · Console, Heart-Lung Machine, Cardiopulmonary Bypass1 recall77 reports
- S5 Roller Pump 150 RP 150LivaNova Deutschland GmbH · CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS1 death120 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page