Skip to content

daVinci Xi 380652Recalls & reported problems

Intuitive Surgical, Inc. · PATIENT SIDE CART, 4-ARM · Class 2

All models of daVinci Xi (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This da Vinci Xi patient side cart is the robotic arm unit of a surgical system. The reports often describe on-screen messages that will not clear before surgery, sometimes forcing arms to be disabled, and some describe unintended movement. FDA records list 5,346 reports, 230 deaths, 858 injuries and 4,258 malfunctions. There are 2 open recalls, about an arm part that may fail and allow uncontrolled movement, and instruments that may deliver energy unintentionally.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
5,316
Deaths reported
235
Injuries reported
854
Recalls naming it
2

Summary

The FDA's MAUDE database contains 5,316 adverse event reports that list this exact device (DI 00886874110720), received between Feb 11, 2019 and Sep 30, 2026 — classified as 235 deaths, 854 injuries, 4,227 malfunctions.

The most frequently reported specific device problem is “Visual Prompts will not Clear” (2,198 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 105 reportsDec 2023: 90 reportsJan 2024: 98 reportsFeb 2024: 27 reportsMar 2024: 71 reportsApr 2024: 66 reportsMay 2024: 69 reportsJun 2024: 83 reportsJul 2024: 70 reportsAug 2024: 79 reportsSep 2024: 64 reportsOct 2024: 107 reportsNov 2024: 105 reportsDec 2024: 56 reportsJan 2025: 90 reportsFeb 2025: 63 reportsMar 2025: 107 reportsApr 2025: 57 reportsMay 2025: 112 reportsJun 2025: 117 reportsJul 2025: 147 reportsAug 2025: 132 reportsSep 2025: 69 reportsOct 2025: 68 reportsNov 2025: 111 reportsDec 2025: 111 reportsJan 2026: 88 reportsFeb 2026: 140 reportsMar 2026: 477 reportsApr 2026: 180 reportsMay 2026: 113 reportsJun 2026: 2 reportsJul 2026: 10 reportsAug 2026: 43 reportsSep 2026: 97 reportsOct 2026: 0 reports202420252026477
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 24, 2023Open, Classified

    Preventative maintenance data was used to identify instrument Arms, part of surgical systems, that allow surgeons to perform surgical tasks, that may be at risk of insertion ballscrew failure, allowing for uncontrolled carriage movement along the insertion axis, which may cause tissue injury.

    Z-1262-2023 · Root cause: Component change control · Recall details · FDA record ↗

  2. Sep 20, 2022Open, Classified

    Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force bipolar is pressed, released with head in surgeon console 4) Blue pedal associated with bipolar is pressed, resulting in force bipolar energy delivery

    Z-0133-2023 · Root cause: Software Design Change · Recall details · FDA record ↗

Most reported problems

Device problems

  • Visual Prompts will not Clear2,198
  • Adverse Event Without Identified Device or Use Problem707
  • Output Problem536
  • Unintended System Motion462
  • No Device Output366
  • Unintended Movement203
  • Physical Resistance/Sticking155
  • Insufficient Device Problem Information112

Patient problems

  • No Clinical Signs, Symptoms or Conditions4,002
  • Insufficient Information456
  • Hemorrhage/Blood Loss/Bleeding312
  • Unspecified Tissue Injury266
  • Cardiac Arrest42
  • Bowel Perforation35
  • Great Vessel Perforation28
  • Perforation of Vessels27

Recent adverse event reports

The 25 most recent of 5,316 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Unintended System Motion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50955020 ↗
  • MalfunctionSep 30, 2026Output Problem

    Patient problems: Insufficient Information

    MAUDE report 50953996 ↗
  • MalfunctionSep 30, 2026Positioning Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50953889 ↗
  • MalfunctionSep 30, 2026Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50953880 ↗
  • MalfunctionSep 30, 2026Visual Prompts will not Clear

    Patient problems: Insufficient Information

    MAUDE report 50953744 ↗
  • MalfunctionSep 30, 2026Visual Prompts will not Clear

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50953500 ↗
  • MalfunctionSep 30, 2026Unintended System Motion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952349 ↗
  • MalfunctionSep 30, 2026Output Problem

    It was reported that prior to starting a da Vinci-assisted surgical procedure, the customer encountered an error 32056 on the Universal Surgical Manipulator 4 (USM4). The Technical Service Engineer (TSE) had the customer emergency power off the Patient Side Cart but the error remained. The TSE advised the customer to disable the USM4, the customer stated that they would confirm with the surgeon that the case could be completed with three Universal Surgical Manipulator (USM)s. The procedure was completed with no reported injury.

    Patient problems: Insufficient Information

    MAUDE report 50942442 ↗
  • MalfunctionSep 30, 2026No Device Output; Visual Prompts will not Clear

    It was reported that the Universal Surgical Manipulator (USM) 1 had a non-recoverable fault 32056 when the system powered on. The customer power cycled the system five times, but the issue remained. The Intuitive Surgical, Inc. (ISI) Technical Service Engineer had the customer hard power cycle the system, but the error returned again. There was no report of patient involvement.ISI contacted the Clinical Territory Associate and obtained the following information: The issue was identified when setting up the Operating Room (OR) in the evening on the 6th to get the system ready for the next day. The two patients (procedures) were cancelled and are being rescheduled. No procedure was going on at the time of the event. The robot was isolated until a Field Service Engineer (FSE) was able to go on site. There was no delay involved, just cancellation of upcoming procedures. There was no injury to any patient. The FSE resolved the issue on the afternoon of the 7th.ISI contacted the Technical Support Engineer and obtained the following information: With error 32056, the customer could have used the system and proceeded with the procedure.

    Patient problems: Insufficient Information

    MAUDE report 50942438 ↗
  • MalfunctionSep 30, 2026Visual Prompts will not Clear

    It was reported that prior to the start of a da Vinci-assisted surgical procedure, the user informed the Technical Support Engineer (TSE) that a repeated recoverable fault 32056 occurred on the Universal Surgical Manipulator (USM3). The user disabled USM3 and continued the procedure using the remaining three USMs without further issues. No known impact or patient consequence was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942433 ↗
  • MalfunctionSep 30, 2026Output Problem

    It was reported that prior to the start of a da Vinci-assisted surgical procedure, the customer encountered an error 32056 on the Universal Surgical Manipulator (USM) 2. The customer power cycled the system several times, but the issue remained. The customer stated that they were moving the case to another room at the time of the call. The procedure was aborted to another da Vinci system with no patient involvement.

    Patient problems: Insufficient Information

    MAUDE report 50942422 ↗
  • MalfunctionSep 30, 2026Visual Prompts will not Clear

    It was reported that prior to the start of a da Vinci-assisted procedure, following anesthesia and prior to port placement, the customer experienced recurrent recoverable errors 1105 and 1162 on Universal Surgical Manipulator (USM) arm#4. Upon reviewing the logs, the Intuitive Surgical Inc. (ISI) Technical Support Engineer (TSE) determined that USM#4 was faulty and required replacement. The TSE guided the caller to disable arm#4 and advised proceeding with the three remaining arms, which the customer indicated would most likely be sufficient for the upcoming procedure. However, the case was ultimately converted to open surgery. No patient injury was reported as a result of the issue or the conversion.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942417 ↗
  • MalfunctionSep 29, 2026No Device Output

    Patient problems: Insufficient Information

    MAUDE report 50939796 ↗
  • MalfunctionSep 29, 2026Output Problem

    Patient problems: Insufficient Information

    MAUDE report 50939338 ↗
  • MalfunctionSep 29, 2026Output Problem

    Patient problems: Insufficient Information

    MAUDE report 50939260 ↗
  • MalfunctionSep 29, 2026Unintended System Motion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50937922 ↗
  • MalfunctionSep 29, 2026Material Separation

    It was reported that outside of da Vinci-assisted surgical procedure, site contacted Intuitive Technical Support Engineer (TSE) to report crack in carriage of Universal Surgical Manipulator (USM) 2.There was no report of patient involvement.

    Patient problems: Insufficient Information

    MAUDE report 50928035 ↗
  • MalfunctionSep 29, 2026Visual Prompts will not Clear

    It was reported that during a da Vinci-assisted surgical procedure, the user informed the Technical Support Engineer (TSE) that repeated recoverable faults occurred on Universal Surgical Manipulator (USM) 3. The user disabled USM 3 and continued with the procedure using the remaining three USMs. The procedure was completed robotically as planned without delay and with no patient harm.On September 10, 2026, Intuitive Surgical (IS) obtained the following additional information regarding this event: Under a related complaint was opened to record the second procedure that took place on the afternoon of (b)(6) 2026, during which the customer experienced the same issue with the system.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50928032 ↗
  • MalfunctionSep 29, 2026Visual Prompts will not Clear

    It was reported that prior to the start of a da Vinci-assisted radical salvage prostatectomy surgical procedure, the site stated they were having recurring recoverable error 32056 on Universal Surgical Manipulators (USM) 1. The nurse restarted multiple times and error returned. The nurse was able to USM 1. The Technical Support Engineer (TSE) reviewed logs and informed that USM 1 was faulty and it has to be replaced. Upon request, the nurse was not able to state if the procedure could take place with the 3 remaining arms. The surgeon took over the call and did not answer to the question. TSE called back reporter to inform about the next step and to collect procedure information. The site informed that they would proceed with 3 Universal Surgical Manipulators (USM). The procedure was completed with no reported injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50928020 ↗
  • MalfunctionSep 29, 2026Visual Prompts will not Clear

    It was reported that prior to the start of a da Vinci-assisted surgical procedure using an Xi system, user informed the Technical Support Engineer (TSE) that there was a continuous error on Universal Surgical Manipulator (USM) 2. Multiple hard power cycles of the Patient Side Cart (PSC) were performed, but the issue persisted. The TSE reviewed the system logs and confirmed the occurrence of the errors. The user converted the procedure to another dV system to continue with the procedure as planned. No known impact or patient consequence was reported.

    Patient problems: Insufficient Information

    MAUDE report 50928019 ↗
  • MalfunctionSep 29, 2026Material Split, Cut or Torn

    It was reported that prior to the start of a da Vinci-assisted surgical procedure, user informed the Technical Support Engineer (TSE) that the Patient Side Cart power cable was damaged and the wall-plug end of the power cable had broken off. There was no report of patient involvement or patient injury.

    Patient problems: Insufficient Information

    MAUDE report 50928015 ↗
  • MalfunctionSep 29, 2026Output Problem

    It was reported that during a da Vinci-assisted surgical procedure, the customer encountered a repeating error 32098 on the Universal Surgical Manipulator 3 (USM3). The customer was able to recover the error by adjusting the USM position but the error would return. The Technical Service Engineer advised the customer to perform a hard power cycle when possible.The procedure was completed with no reported injury.Intuitive Surgical, Inc. (ISI) followed up with the initial reporter and obtained the following additional information: The customer did disabled the arm 3 and completed the case with the remaining three arms and a laparoscopic instrument in the arm 3 port. The issue cased a 30 minute surgical delay.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50928009 ↗
  • MalfunctionSep 29, 2026Visual Prompts will not Clear

    It was reported that prior to the start of a da Vinci-assisted general inguinal hernia surgical procedure, a recoverable fault 32056 had been reported and error 32056 on Universal Surgical Manipulator (USM) arm 4 was observed. Technical Service Engineer (TSE) confirmed the error on the system logs. The customer performed an Emergency Power Off (EPO) of the Patient Side Cart (PSC); however, the issue persisted. The procedure was continuing with USM arm disabled with no report of patient involvement.

    Patient problems: Insufficient Information

    MAUDE report 50928007 ↗
  • MalfunctionSep 29, 2026Output Problem

    It was reported that prior to start of da Vinci-assisted surgical procedure, site contacted Intuitive Technical Support Engineer (TSE) to report error 32056 on Universal Surgical Manipulator (USM) 3 and could not be cleared after restarting the system. I had the customer perform two hard power cycles, but the issue did not recover. Disabling the affected arm was difficult. The procedure was converted to laparoscopic with no reported injury.

    Patient problems: Insufficient Information

    MAUDE report 50927998 ↗
  • MalfunctionSep 28, 2026Visual Prompts will not Clear

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50926065 ↗

Other System,surgical,computer controlled instrument devices

All System,surgical,computer controlled instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page