daVinci Xi 381121Recalls & reported problems
Intuitive Surgical, Inc. · VISION SIDE SYSTEM · Class 2
All models of daVinci Xi (4)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This da Vinci Xi vision side system is part of Intuitive's robotic surgical system. Its 4,368 adverse event reports mostly describe energy output problems and generator error codes, and some describe burns and bleeding. The reports include 2 deaths and 264 injuries. 1 recall names this device: started in September 2022 and still open, it covers energy being delivered unintentionally when certain bipolar instruments and generator connections are combined.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 4,342
- Deaths reported
- 2
- Injuries reported
- 268
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4,342 adverse event reports that list this exact device (DI 00886874110898), received between Dec 15, 2016 and Sep 30, 2026 — classified as 2 deaths, 268 injuries, 4,069 malfunctions.
The most frequently reported specific device problem is “Output Problem” (1,929 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Sep 20, 2022Class IIOpen, Classified
Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force bipolar is pressed, released with head in surgeon console 4) Blue pedal associated with bipolar is pressed, resulting in force bipolar energy delivery
Z-0133-2023 · Root cause: Software Design Change · Recall details · FDA record ↗
Most reported problems
Device problems
- Output Problem1,929
- Visual Prompts will not Clear917
- Energy Output Problem587
- Failure to Deliver Energy297
- Grounding Malfunction148
- Power Problem111
- No Device Output77
- Continuous Firing58
Patient problems
- No Clinical Signs, Symptoms or Conditions3,821
- Insufficient Information314
- Unspecified Tissue Injury108
- Hemorrhage/Blood Loss/Bleeding41
- Burn(s)38
- No Known Impact Or Consequence To Patient17
- Bowel Perforation4
- Full thickness (Third Degree) Burn3
Recent adverse event reports
The 25 most recent of 4,342 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Visual Prompts will not Clear
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50954542 ↗MalfunctionSep 30, 2026Material Split, Cut or Torn
Patient problems: Insufficient Information
MAUDE report 50954384 ↗MalfunctionSep 30, 2026Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50953600 ↗MalfunctionSep 30, 2026Output Problem
It was reported that Universal Surgical Manipulator (USM) 2 faulted with error 32056 while the site's Clinical Sales Representative (CSR) was troubleshooting video issues. The Intuitive Surgical Inc Technical Support Engineer (TSE) reviewed the system logs and confirmed the reported error. The TSE had the CSR perform a hard reboot, with no change. The caller stated that the site has several cases scheduled for the following day that they would need all 4 USMs for. There were no reports of patient involvement.
Patient problems: Insufficient Information
MAUDE report 50942414 ↗MalfunctionSep 29, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50939848 ↗MalfunctionSep 29, 2026Visual Prompts will not Clear
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50939638 ↗MalfunctionSep 29, 2026Output Problem
It was reported that prior to starting a da Vinci-assisted surgical procedure, an error occurred during startup and did not clear after a restart. The troubleshooting first confirmed the initial code as a different error code, and it was noted that an extension cable had been in use. The connection was then changed and the unit was restarted, but the error still did not clear. An alternate device was available onsite but had not been integrated, so the integration method was explained and an external pedal option was discussed, though those options were not accepted. After the error, integrated mode had been used and ¿Not Available¿ had displayed during use, making energy activation unavailable. The installed instruments across all arms were reviewed, and it was considered whether one instrument might have been detected as a second bipolar instrument. Instructions were provided to discontinue integrated operation and set up the external pedal, after which energy output had been available via the pedal even though ¿Not Available¿ continued to display. The customer was to test further and recontact support if energy output remained unavailable.The procedure was completing as planned
Patient problems: Insufficient Information
MAUDE report 50928040 ↗MalfunctionSep 29, 2026Output Problem
It was reported that during a training/demo of the da Vinci system, the external bipolar pedal had been sticking and had caused a system error. Technical Service Engineer (TSE) guided the caller to disconnect the pedal from the energy device, which cleared the error. There was no report of patient involvement.
Patient problems: Insufficient Information
MAUDE report 50928005 ↗MalfunctionSep 28, 2026No Device Output
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50926057 ↗MalfunctionSep 28, 2026Output Problem
Patient problems: Insufficient Information
MAUDE report 50925981 ↗MalfunctionSep 28, 2026Output Problem
Patient problems: Insufficient Information
MAUDE report 50924917 ↗InjurySep 28, 2026Visual Prompts will not Clear
It was reported that during a da Vinci-assisted surgical procedure, the customer encountered a system error while prepping for the case. The customer restarted the system and continued with set up for the case. However, the customer called an Intuitive Surgical, Inc. (ISI) Technical Support Engineer (TSE) back and reported the error(s) reoccurred. Therefore, the customer elected to convert the procedure to open surgery.\\Intuitive Surgical, Inc. (ISI) has attempted to gather additional information regarding the reported event; however, no response has been received.
Patient problems: Unspecified Tissue Injury
MAUDE report 50918155 ↗InjurySep 28, 2026Energy Output Problem
It was reported that during a da Vinci-assisted right hemicolectomy surgical procedure, the Vessel Sealer Extend (VSE) instrument did not work as expected with the E-200 energy generator. The sealing did not complete as usual and, after repeated attempts, the instrument jaws stuck to tissue when opened. An Intuitive Surgical (IS) Technical Support Engineer (TSE) was contacted for troubleshooting support. It was also reported that this issue did not occur when other systems were used with different energy generators. The TSE reviewed the event information and noted a very large number of occurrences of a specific error code during the same procedure, which was considered not realistic to have been caused by another instrument connected to the second bipolar port during vessel sealing.\\The Vessel Sealer instrument was not sealing and gave an error indicating a non-complete sealing cycle. Tissue was adhered to the instrument which caused tissue damage and bleeding. Tissue was never cut that was not fully sealed and the instrument was noted to work ¿sometimes¿ with audible tones. \\The procedure was converted to laparoscopic with no additional ports placed. Information surrounding how
Patient problems: Hemorrhage/Blood Loss/Bleeding; Unspecified Tissue Injury
MAUDE report 50915536 ↗MalfunctionSep 27, 2026Power Problem
Patient problems: Insufficient Information
MAUDE report 50909579 ↗MalfunctionSep 27, 2026Visual Prompts will not Clear; Output Problem
Patient problems: Insufficient Information
MAUDE report 50909578 ↗MalfunctionSep 25, 2026Energy Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50905258 ↗MalfunctionSep 25, 2026Output Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50905031 ↗MalfunctionSep 25, 2026Output Problem
Patient problems: Insufficient Information
MAUDE report 50904524 ↗MalfunctionSep 25, 2026Visual Prompts will not Clear
It was reported that during a da Vinci-assisted surgical procedure that non recoverable fault 86 occurred. The customer hard power cycled the system and performed an emergency power off, but the error reoccurred. The customer converted the procedure to laparoscopic. There were no reports of patient injury.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50900751 ↗MalfunctionSep 25, 2026Output Problem
It was reported that during da Vinci-assisted surgical procedure, ERBE had C00 error. No further information was available.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50893632 ↗MalfunctionSep 24, 2026Output Problem
Patient problems: Insufficient Information
MAUDE report 50891432 ↗MalfunctionSep 24, 2026Continuous Firing
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50890920 ↗MalfunctionSep 24, 2026Output Problem
Patient problems: Insufficient Information
MAUDE report 50890919 ↗MalfunctionSep 24, 2026Output Problem
Patient problems: Insufficient Information
MAUDE report 50890797 ↗MalfunctionSep 24, 2026Output Problem
Patient problems: Insufficient Information
MAUDE report 50890714 ↗
FDA clearance and classification
The FDA regulates product code NAY (system,surgical,computer controlled instrument) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the daVinci Xi 381121
Has the daVinci Xi 381121 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0133-2023 (Class II), began on Sep 20, 2022: Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3)… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the daVinci Xi 381121?
The FDA's MAUDE database lists 4,342 adverse event reports for this device, received between Dec 15, 2016 and Sep 30, 2026. In those reports, the most common specific device problems are “Output Problem” (1,929), “Visual Prompts will not Clear” (917) and “Energy Output Problem” (587); the most common patient problems are “Unspecified Tissue Injury” (108), “Hemorrhage/Blood Loss/Bleeding” (41) and “Burn(s)” (38). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the daVinci Xi 381121?
2 reports are classified as deaths, 268 as injuries and 4,069 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the daVinci Xi 381121, and is it still sold?
The FDA's device database lists Intuitive Surgical, Inc. as the company responsible for it (brand daVinci Xi, model 381121). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other System,surgical,computer controlled instrument devices
- EndoWrist 471205Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall14,308 reports
- EndoWrist 470179Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2 deaths10,204 reports
- EndoWrist 471093Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall8,630 reports
- EndoWrist 471172Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death6,325 reports
- daVinci Xi 380652Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls233 deaths5,369 reports
- EndoWrist 471405Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death3,802 reports
- EndoWrist 470327Intuitive Surgical, Inc. · System,surgical,computer controlled instrument2 recalls2,281 reports
- EndoWrist 471049Intuitive Surgical, Inc. · System, surgical, computer controlled instrument1 recall1 death2,285 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page